Inogen One® G2 System Specifications Chapter 8
English
NOTE At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects, and people.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the [ME EQUIPMENT or ME
SYSTEM] requires continued
operation during power
mains interruptions, it is
recommended that the
[ME EQUIPMENT or ME
SYSTEM] be powered from
an uninterruptible power
supply or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Compliance
Level
IEC 60601 Test LevelImmunity Test Electromagnetic
Environment - Guidance
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
(continued):
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-4-8
<5 % U
T
(>95 % dip
in U
T
) for 0.5 cycle
40 % U
T
(60 % dip
in U
T
) for 5 cycles
70 % U
T
(30 % dip
in U
T
) for 25 cycles
<5 % U
T
(>95 % dip
in U
T
) for 5 sec
<5 % U
T
(>95 % dip
in U
T
) for 0.5 cycle
40 % U
T
(60 % dip
in U
T
) for 5 cycles
70 % U
T
(30 % dip
in U
T
) for 25 cycles
<5 % U
T
(>95 % dip
in U
T
) for 5 sec
3 A/m 3 A/m