User's Manual

D-2 Appendix Installation Manual
Tags (no detection)
Tag is not detected Use a TAD to verify Tag is turned on and
functioning properly
Tag (turned on, functioning) is not detected See “No Tx indicator” section
See “No Rx” section
False Tag Detection
(nuisance alarms)
Is another zone setting this zone into alarm?
Are patient rooms adjacent to monitored zone?
Implement Stagger Tuning (page 4-6)
Reduce Tx wand antenna sensitivity
(page 4-7)
Reduce Receiver sensitivity
(page 5-6)
Keypad
Does not reset
Verify code
Was incorrect code entered 3 times?
If so, Keypad will lockout user for 90 seconds
(
Green LED blinks during this time)
Verify proper wiring and voltage
Replace Keypad
(from another zone or a spare)
Tx Supervise alarming
(LED14, steady Red)
Are metal carts in front of antennas? Remove metal carts from area
Is antenna damaged? Replace antenna (from another zone or a spare)
Is cable broken or pinched? Check for broken or pinched cable
Is the antenna(s) tuned properly? See Tuning procedure (page 4-7)
Is Tx Supervise threshold set too high? See Tuning procedure (page 4-7)