User's Manual
• 54 • • 55 •
Troubleshooting
The test result is lower
than the following.
Glucose : less than
20mg/dL
TC : less than 100mg/dL
TG : less than 70mg/dL
HDL : less than 25mg/dL
Please retest with a new test strip.
If the problem persists, please contact
your local representative or customer
support.
Lipid profile : The
ambient temperature is
over 86ºF(30ºC).
Glucose : The ambient
temperature is over
104ºF(40ºC)
Place the meter at the temperature
between 64~86ºF(18~30ºC) for lipid
profile test and 50~104ºF(10~40ºC) for
glucose test for more than 10 minutes and
test again
Lipid profile : The
ambient temperature is
less than 64ºF(18ºC)
Glucose : The ambient
temperature is less than
50ºF(10ºC)
Place the meter at the temperature
between 64~86ºF(18~30ºC) for lipid
profile test and 50~104ºF(10~40ºC) for
glucose test for more than 10 minutes and
test again.
This message appears
with 3 times of beep
sound if the mode button
is pressed while no test
strip of lipid profile is
inserted.
Please make sure to insert a test
strip of lipid profile when performing
the test.
Please recognize the code and retest
(refer to page 21)
This message appears
with 3 times of beep
sound with ‘CODE’ and
‘Er’ on the display are
none memory code
and/or unrecognized
strip used.
Message Probable cause Action required
Product Specifications
Inconsistent or unexpected test results
If you continue to get unexpected results, check your system with control
solution.
If you experience symptoms that are not consistent with your lipid and glucose
results, first be sure you have followed all instructions in this Instructions for
use, then, contact Infopia Customer Service. Never ignore symptoms or make
significant changes to the control program.
Product Specication
Devision Content
Sample type
Measuring principle
Testing Types
Testing method
Testing range
Testing time
Calibration
Sample volume
Operating Humidity
Operating Temperature
Storage Humidity
Storage Temperature
(meter, Strip)
HCT Range
Capillary whole blood, Serum, Venous blood
Spectroscopy
Total cholesterol HDL-cholesterol
Triglyceride LDL-cholesterol
Enzymatic-colorimetric method
Total cholesterol : 100~400mg/dL,
Triglyceride : 70~600mg/dL,
HDL-cholesterol : 25~80mg/dL
About 2mins
Plasma_Equivalent
7uL
10 ~ 90 %
64 ~ 86ºF(18 ~ 30ºC)
10 ~ 90 %
36 ~ 86ºF(2 ~ 30ºC)
30 ~ 55%
Lipid Profile
Test