User's Manual

INFOPIA CO., LTD Page 8 UM-M (Rev.0)
1. Unscrew the lancing
device cap.
2. Insert a sterile lancet into
the lancing device.
3. Twist the protective
cover off. Do not
discard.
4. Put the lancing cap
back and twist it
tightly.
5. To adjust the depth
setting:
Use strength level 1-2 for
soft skin, 3 for average
skin, 4-5 for thick or
calloused skin.
6. Pull the end of the
lancet device back.
7. To prick your finger,
push the center
button.
CAUTION
1. If the blood smears or runs, do not use such blood
sample. Dry the area and gently squeeze another drop
or puncture a new area of the finger.
Do not share reusable lancing device with anyone
Step 3
IGM-1001D test meter only requires 0.5 ul sample volume.
Place the edge of the test strip to the drop of blood.
The blood will automatically be ‘drawn’ into the test strip
channel.
When your sample blood is enough, your meter will
automatically count down
CAUTION
1. If the countdown does not start, do not add more blood
to the glucose test strip. Discard the glucose test strip
and re-start testing.
2. If you do not conduct the test within 3 minutes, the
glucose test meter will automatically power off to save
battery life. In this case, test procedure should start
again from the beginning.
3. You may get an inaccurate result if the blood sample is
not completely filled.
4. You must inject the blood sample until beep sounds.
Step 4
After the beeping sound, the test will begin automatically
and your results will appear in 5 seconds. It should begin
counting down from 5 to 1 second on the LCD display
window. The LCD window will display the result of your
blood glucose level. Temperature and time.
After results displayed, you choose to use the
activity/meal flags ( ), press and
release or button to select the activity, then
press and release the power button