4. Dip the test strip into the control
solution.
Results will appear in 5 seconds.
Compare the result to the range
printed on the test strip vial, the
result should fall within that range
CAUTION
1. Control solutions is for in-vitro diagnostic use only. This
means that it is only used for testing outside the body.
2. Control solution is exclusively made for the IGM-1001D
Blood Glucose Monitoring System. Do not use other brand
control solutions made by other manufacturers.
3. Check the expiration date on the vial. Do not use, if
expired.
4. Once opened, use it for the next three months only.
Record the discard date on the control solution bottle.
Discard after three months.
5. Do not swallow. This is NOT for human or animal
consumption.
6. The control solution bottle icon contains letter ‘C,’ so as
not to be confused with the medication bottle.
This will allow you to understand differences between a
control solution test and an actual blood test.
For example only :
If you are using the normal control solution, according to this
particular vial of strips, your meter should show a number
between 80 and 120.
Please note that the control values shown in the picture are not
control ranges.
NOTE:
- It is recommended that glucose control solution is stored at room
temperature 20~25℃ (68~77℉) before testing.
- Check the expiration date before performing a glucose control
solution test.
- Do not use if it expired. Once opened, the control solution expires
after 3 months.
If your control solution test falls out of range, please follow
the next steps before contacting customer support:
Check the expiration dates on all the products you are
using. Be sure you are using IGM-1001D control solution.
Try another control solution test.
If this test falls out of range, try another control solution
test with a new unopened bottle of strips.
After following the appropriate steps and the control test
still falls out of range, do not perform a glucose test.
Please call customer support toll: 82-31-460-0300
Follow-up Action
1. Check your meter and strip with the glucose control
solution.
2. Perform your test again.
CAUTION
1. If your glucose control solution falls out of range, do not
perform a glucose test and please contact your local
distributor.
2. The glucose control solution’s measurement range has
nothing to do with individual’s blood glucose level, as its
own purpose is to check the status of strip and meter.
Discard the used control solution and test strips carefully,
according to the local regulation.
Performing Your Test
Testing Blood - Preparation
Lancing device is a convenient medical tool for collecting
capillary blood samples for glucose monitoring or other
tests which require one or two drops of blood.
Being a more advanced lancing device, it provides
ultimate safety and comfort for obtaining blood samples.
An adjustable tip offers 5 different strength levels of skin
penetration for individual user’s comfort, while a lancet
ejector enables the safe disposal of the used lancet. As
an added bonus, this also comes with a transparent tip
for AST (Alternative Site Testing).
CAUTION
1. Never use a lancet that has been used by someone else. This
could lead to contamination.
2. If the lancing device is to be used by another person, the unit
must use a new tip or properly disinfected
3. Do not leave the lancet in the device after use.
4. A new lancet must only be placed into the lancing device
directly before testing.
5. This device has many small parts and could be a choking
hazard for children if swallowed.
6. In the case of hospital use, hospitals need to observe their
own infection control protocols in order to avoid any
contamination.
7. Always dispose of the used lancet in a biohazard garbage
container.