User's Manual

INFOPIA CO., LTD Page 4 UM-M (Rev.0)
CAUTION
1. Store the IGM-1001D
test strip vials in a cool, dry place.
Keep out of direct sunlight. Do not freeze.
2. Store test strips in their original vial only. Do not mix the
test strips in new vials or in any other container.
3. Immediately replace the vial cap and close tightly after
removing any test strips from the vial.
4. Make a note of the expiry date, which is three months
from the date you first open a new vial of strips. Throw
IGM-1001D
test strips and vial away after the expiry date.
5. Do not use the test strips after the expiration date printed
on the package or vial, since it may cause inaccurate
results.
6. IGM-1001D
test strips should be single use only. DO
NOT RE-USE.
7. Do not test at temperatures below 10(33.8°F) or above
40(104°F).
8. Do not test in the condition that humidity is below 10% or
above 90%.
9. Do not bend, cut, or alter the test strip.
10. Avoid getting dirt, food, and water on the test strip with
wet hands.
11. Avoid getting dirt, food, and water on the color-coding
label (backside of test strip)
12. Refer to additional information in the IGM-1001D
test
strip package.
Before Testing
WARNING
To reduce the chance of infection:
1. Lancing device and lancets should NOT be shared with
others.
2. Always use a new lancet and a new blood glucose test
strip.
3. Practice using the lancing device and become
accustomed with its use.
4. Lancets and blood glucose test strips should be single
use only.
5. Wash your hands in warm clean running water using
soap before testing.
6. Avoid getting hand lotion, oils, dirt or debris on the
lancets or on the lancing device.
7. Dry your hands completely before testing.
Setting Your Meter
The IGM-1001D meter has a wide variety of functions to
choose from. From the setup mode you have the ability to
turn activity/meal flags, set the date/time, designate three
unique averages and set up to five daily alarms.
Power On
Press for 3
seconds at least
User Activity Option
After pressing and
releasing either ▲ or
buttons to turn the
User Activity Option
on/off, confirm with
Year
After pressing and
releasing either▲ or
buttons to set the
year, confirm with
Date / Time
After pressing and
releasing either▲ or
buttons to set the
date and time,
confirm each with
Number of days for
average
After pressing and
releasing either▲ or
buttons to set the
number of days for
average calculation (3
different possible),
confirm with
Alarm on/off
After pressing and
releasing either▲ or
buttons to turn the
alarm on/off, confirm
with
Alarm time
After pressing and
releasing either ▲ or
buttons to set the
alarm hour and
minute, confirm with
Alarm
After programming
the number of desired
alarms (5 alarms
possible), confirm
with
CAUTION
1. VERY IMPORTANT: Set the right test unit.
Your IGM-1001D Meter is pre-set to mg/dL as test unit. A
wrong test unit will read to wrong test result display.
Please consult your healthcare professional if you’re not
sure which test unit to use. Without setting the date
properly, the average glucose level and the results in
memory will not show proper values. It is recommended
to set the glucose test meter before use and also when
new batteries are installed.
2. You cannot test your blood glucose in the setting mode.
3. To turn the meter off during the setup, press for at
least 3 seconds.