I N T E R F A C E D E S C R I P T I O N iins70e1-c (1005) Catalog No. 550-000, 550-001 550-100, 550-101 550-500, 550-501 from software version V 4.4 (UL1000) V 4.4 (UL1000 Fab) V 4.
0-2 iins70eIVZ.
1 Interface Description 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.8.1 1.8.2 Preface Interface Connecting Cable Command Format Error messages Parameter Examples Command list Calibration via RS232 Internal Automatical Calibration External Calibration 1-1 1-1 1-1 1-1 1-3 1-3 1-3 1-4 1-8 1-8 1-8 iins70eIVZ.
0-4 Content iins70eIVZ.
Interface Description 1.1 Preface Operating Manual 1 The RS232 porotocol uses an ASCII format: so every leak detector can be controlled with a simple terminal programm. Choose the interface porotocol „ASCII“ to connect the leak detector with the computer via RS232 interface. In case the leak detector shall also be controlled via computer (START, STOP, ZERO, etc.), set the “control location” at the leak detector to „RS232“ or „local and RS232“.
Some commands can be used as queries, some can be used to set menu parameter and some can be used for both. A query is marked by a „?“ (ASCII code 63dec) after the command, for setting data the command has to be followed by the new value to be set. Boolean 0 / 1 or OFF / ON Numeric representation format: integer, real (15.6) or exponential (4.5E-7) Format: [space] [sign] [ddd] [.] [ddd] [e[sign]ddd] (d: digit) Notice Always use a point as the decimal marker.
Error messages Operating Manual 1.4 1.
Command list *IDN :CRC :DEVice :VERsion :SERial :TURBO :MC68 :IOversion :GBversion :VDversion :DIP1 :DIP2 :TURBO *STATus identification check sum name of instrument (UL1000, UL1000Fab, UL5000) software version serial-Number software version TC600 hardware identification MC68 hardware identification IO-Karte hardware identification control panel hardware identification mother board MC68 DipSwitch 1 MC68 DipSwitch 2 software version frequency converter :TESTLog mode of calibration routine INIT, ACCL, STBY
*MEAS Operating Manual :P1 :P2 :P3 :OFFset :IMess :IFilter :UNV :UVV :MIAP :MIKP :MISP :MIAKP :VALVE :U24EXT :U24FB :U24FAN :TEMPeratur inlet pressure foreline pressure inlet pressure (UL5000) offset current [A] current raw values [A] current filtered [A] amplifier voltage [V] preamplifier voltage [V] anode potential [V] cathode potential [V] suppressor potential [V] anod-/cathode potential [V] valve voltage [V] external voltage [V] voltage remote contol [V] fan voltage [V] :Amplifier :Electronic preamp
:MINVOLume :MODE :PARTCOUNT :PARTNO :PEVACGROSS :PGROSSFINE :PROTection :CONTamination :CONTLimit :EVACtime :PARTicle :PMAX :PMIN :PURGe :RECorder :LINK1_2 / 12 :LINK3_4 / 34 :RS232 :SUPPRession :TESTINGTime :TESTINGTIMEmax :TESTINGTIMEMIN :TIMEAXIS :TMPVENT :TRIGger1 :TRIGger2 :UNIT mass 2 mass 3 mass 4 minimum audio volume (0 ...
:SNIFf :M2 :M3 :M4 iins70e1.
1.8 Calibration via RS232 1.8.1 Internal Automatical Calibration When the leak detector is in STAND-BY mode the command „*CAL“ evokes an internal autmatical calibration. The process is completely automatical. Via „STATus?“ the computer can realise if the calibration routine is finished. 1.8.
Interface Description 1-9 iins70e1.
INFICON GmbH, Bonner Strasse 498, D-50968 Cologne, Germany Phone: +49 (0)221 347-40 Fax: +49 (0)221 347-41429 E-mail: leakdetection@inficon.com UNITED STATES TAIWAN JAPAN KOREA SINGAPORE GERMANY FRANCE UNITED KINGDOM Vi s i t o ur w eb s i t e f o r c on ta c t i nf o rma t i on an d o t he r s al es of f i c e s w or l dwi de . w w w. i n fi c o n .