User guide
4 - 9
PN 074-156N
Front Load Single and Dual Sensors Operating Manual
4.3.5 Lubricating the Shutter Module
The shutter module should be lubricated approximately every 2000 strokes at
areas specified as shown in Figure 4-8. Failure to lubricate the shutter module may
significantly reduce life of operation or cause assembly to become inoperative.
For lubrication, use molybdenum disulfide (PN 750-191-G1), provided with each
shuttered sensor, or use Fomblin® E25 (perfluorinated polyether), if appropriate for
the process.
Figure 4-8 Lubrication guide
1
2
3
NOTES:
1 - Apply molybdenum disulde to the coupling stem when
extended of 1.
2 - Apply molybdenum disulde to the shaft hole of 2.
3 - Apply molybdenum disulde to the pin and hole of 3.