Table of Contents About this manual ........................................................................................................................................................ 5 Target groups ................................................................................................................................................................. 5 Other associated documents ....................................................................................................................
Table of Contents Interface Description Ecotec E3000, kins22e1-j, 1401
1 About this manual 1.1 Target groups This Interface Description is intended for the owner and for technically qualified personnel with experience in leak detection technology and integration of leak detection devices in leak detection systems. In addition, the installation and use of the unit require knowledge of electronic interfaces. 1.
1 About this manual Interface Description Ecotec E3000, kins22e1-j, 1401
2 I/O Port The I/O part allows communication with external equipment via analog data (programmable logic control - PLC). Through this connection some functions of the Ecotec E3000 can be controlled externally or measurement data or the Ecotec E3000 status may be communicated to external equipment. Through relay changeover contacts the trigger levels as well as the operating mode (Ready) of the Ecotec E3000 may be monitored. 2.1 Connecting the I/O Port Hazardous voltage.
Table 1: I/O Port - pin assignment Pin Signal Notes 2, 3, 18, 19, 21 Ground Ground connectors of the control voltages. 6 +24 V Common output for supplying the PLC inputs and outputs, internally protected with fuse F4 7 Sleep (default) Change from low to high to activate. Change from high to low to deactivate (wake up).
Gas select, gas a, gas b Through the control inputs Gas … (pin 9, 13 and 25) it is possible to select a gas loaded in the active gas memory (one gas out of four). The Ecotec E3000 will then operate in the single gas mode. With an edge triggered signal from LOW to HIGH at Pin 25, the selected gas at Pin 13 and 9 becomes valid. Table 2: Gas select 2.1.
Technical Data relay outputs Material damage if load is high. Wrong input voltage or current can destroy the unit. ► Pay attention to the technical data of the relay outputs: Maximum load rating is 60 V DC / 25 V AC and 1 A per relay (resistive load). Relay output Active = Normally open contact (NO) closed Trigger If the preset trigger level is exceeded, relay contact NO will close and the relay contact NC will open respectively.
LOG The voltage range spans from 0 V to 10 V. Beginning at 1 V, each leak rate decade is spread over 2 V, i.e.: 1.0 … < 3 V: 1st decade 3.0 … < 5 V: 2nd decade 5.0 … < 7 V: 3rd decade 7.0 … < 9 V: 4th decade Thus a leak rate range of 4 decades will be output by way of an analogue signal. The four decades are set through the value and the unit of measurement for the trigger level.
Recorder output gas Here one gas from the four gases is defined which will be output as the analog signal through channel 1. Selecting one of the numbers 1 … 4 outputs the corresponding gas 1 … 4 on the display as the analog signal. Channel 2 (pin 14 of the I/O port) indicates the gas number by way of a voltage, i.e. a voltage between 1 … 4 V will be present accordingly.
Table 6: PLC Output, pin - command assignment Pin Command Pin Command 4 READY 16 TRIGGER 5 not used 17 ERROR Table 7: Relays outputs Pin Type of contact Relay Output (default) 10 Currentless closed (NC) TRIGGER 22 Common 23 Currentless opend (NO) 24 Currentless closed (NC) 11 Common 12 Currentless opend (NO) READY To change these settings select the appropriate pin with the UP and DOWN arrows on the left side of the display and afterwards select the desired command from the lis
Leak rate Cal (Input) Ready (Output) Cal stable (Output) Measurement helium Measurement air Fig. 5: Calibration timing interval A calibration process is started by setting the calibration input to HIGH after the sniffer probe tip has been placed in front of a test leak. The ready signal will go to HIGH after the calibration process has started. If a calibration is started within the first 20 min after power on a warning is issued.
3 RS-232 interface The Ecotec E3000 is equipped with a RS-232 interface which is located on the rear right side of the instrument. This interface is of the DCE type (Data Communications Equipment) and allows the connection of a PC for monitoring and data logging. The connection is provided through a commercially available Sub-D plug. The RS-232 protocol can be set to “ASCII”, “Diagnostics”, “Printer auto” or “Printer manual”.
3.2 Interface Commands 3.2.1 Command Format In ASCII mode any command starts with « * (ASCII code 42 dec)» and is finished with the end sign selected (e.g. CR). There is no differentiation between upper and lower case (exception: gas names and names of I•Guide programs). A blank is required between the command and the parameter, no other blanks are allowed (except names of I•Guide programs). There is a short and an extended form of the command.
3.2.2 List of Commands The following table lists all commands available in the ASCII mode.
Command Command word1 word2 18 3 RS-232 interface Command word3 Statement Read/Set :ECOcheck ECO-Check (ON, OFF) R/S :ENDsign end sign (CR, LF, CR LF) R/S :FILament filament (A, B) R/S :FILTer filter type (AUTO, FIXed, I•Filter) R/S :FLOWHigh flow limit high [sccm] (160 … 999) R/S :FLOWLow flow limit low [sccm] (0 … 160) R/S :INVERSE Reverse colour of the main unit display R/S :LANGuage language (ENGlish, DEUtch, FRAncais, ITALiano, R/S PORtugese, ESPanol, Japanese in KATak
Table 11: List of commands, ASCII (cont.
Table 11: List of commands, ASCII (cont.) Command Command word1 word2 Command word3 Statement Read/Set :MODE Mode enable: on, 1, ENAble disable: off, 0, DISable R/S :NAME Gas name (e.g. R134a) R/S :PERcent Concentration [%] R/S :SEARch search level (5 … 100) [%] R/S :TLRate Leakrate external leak (4.1. [TLUnit]) (proper setting of molecular mass required) R/S :TLUnit Unit external leak (g/a, oz/yr, ppm, mbar*l/s, Pa*m3/s, atm*cc/s, Torr*l/s) R/S :TRIgger Trigger level (2.
Table 11: List of commands, ASCII (cont.
Table 11: List of commands, ASCII (cont.) Command word3 Statement Read/Set :NAME name (max. 6 characters) R/S :NR :NRA gas number (1 … 4) R/S :NRB Gas number B (1 … 4) R/S :POInts Number of points (0, 1, … 20) R/S :TRIGger :TRIGA Global trigger A [in selected unit] (unit read only, determined by gas number) R/S :TRIGB :TRIGgerB Global trigger gas B [selected unit] (unit read only, determined by gas number B) R/S :WAITtime Wait time [s] (1.0, 1.1, … 25.
Table 11: List of commands, ASCII (cont.
Command Command word1 word2 Statement Read/Set [:SNkey] State Sniffer key Byte 1 left Byte 2 right R [:LEAK] ECO-Check test leak R [:Gas] gas type (e.g.
Table 11: List of commands, ASCII (cont.
3.2.3 Error Messages The following error messages may be received from the Ecotec E3000: Table 12: Error messages 3.
Diagnostics The diagnostics protocol is a binary protocol used for analysis in quality control and trouble shooting during manufacturing and service. Default setting: ASCII Printer manual The Printer manual protocol allows to send leak rates on pressing the right sniffer probe button. One line will be send for each currently enabled gas. When in any of the two printer modes, no ASCII commands must be sent to the Ecotec E3000 as this will interrupt the printer mode.
3.4 Examples of programming 3.4.
Table 13: Commands and responses measurement mode (cont.) Command sent to Response received from Meaning Ecotec E3000 Ecotec E3000 *status? (CR) Requests the status of the Ecotec E3000. MEAS (CR) 3.4.2 Ecotec E3000 is in measurement mode. Performing an external calibration Gas settings and interface settings as above. With the following chain of command an external calibration of gas 1 (R134a) may be controlled via RS-232.
Table 14: Commands and responses external calibration (cont.) Command sent to Response received from Meaning Ecotec E3000 Ecotec E3000 *cal:status? (CR) Requests the status of the calibration. LEAK STABLE, CONFIRM (CR) Status: waiting for acknowledgement that signal from test leak is stable. *cal:read? (CR) Requests leak rate signal reading. 8.2638e-14 (CR) .......... Signal from test leak is 8.2638e-14.
Table 14: Commands and responses external calibration (cont.) Command sent to Response received from Meaning Ecotec E3000 Ecotec E3000 CAL (CR) Status: still in calibration mode (calibration process not finished yet). .......... *status? (CR) Requests the status of the Ecotec E3000. MEAS (CR) Ecotec E3000 is in measurement mode. Interface Description Ecotec E3000, kins22e1-j, 1401 The calibration process may be cancelled at any time with “*cal:esc (CR)”.
3 RS-232 interface Interface Description Ecotec E3000, kins22e1-j, 1401
4 Trouble Shooting Table 15: Trouble Shooting Error Possible Reason Solution No characters are received via the interface/the Ecotec E3000 does not answer. Wrong cable Please use a 1:1 cable, (NO null-modem cable, also called cross-over cable!). Problems with flow control Deactivate flow control in PC/PLC or use cable according to the wiring diagram in Section 2. Wrong COM-Port used at PC. Select correct COM-Port.
4 Trouble Shooting Interface Description Ecotec E3000, kins22e1-j, 1401
Interface Description Ecotec E3000, kins22e1-j, 1401
INFICON GmbH, Bonner Strasse 498, D-50968 Cologne, Germany Phone: +49 (0)221 347-40Fax: +49 (0)221 347-41429E-mail: leakdetection@inficon.com UNITED STATES TAIWAN JAPAN KOREA SINGAPORE GERMANY FRANCE UNITED KINGDOM HONG KONG Vis it our web si te fo r co nta c t in fo r ma t io n a nd o t her s a l e s of f ices wo r ld wi de. w w w.i nf ico n.