Instruction Manual

28
tina88e1-b (2014-09)
4.2 Configuring the network interface
The network interface can be configured via the wired or the
wireless network.
4.2.1 Changing IP Address of the Wired Network
The standard IP address of the wired network (default
192.169.0.248
5)
) can be changed via
Web interface "Configuration" ( [4])
RS232 ASCII interface with command
IPL [IP address]|[Subnet mask] ( [2]):
e.g. IPL 10.0.0.200|255.255.255.0
Ethernet interface with REST services
connect Cube with a computer via Ethernet cable ( 12)
set the Ethernet interface of the computer to any IP
address in the subnet 192.168.0.248|255.255.255.0
(e.g. 192.168.0.247)
open a web browser (e.g. Mozilla Firefox, version >20, or
Windows Internet Explorer, version 10.0)
set the new IP address with the REST services command
http://192.168.0.248:8087/1/cmd/IPL%20
[IP address]|[Subnet mask]
An automatic reset of the gauge is performed, if the IP
address is successfully changed.
4.2.2 Activating the Wireless Network
The wireless network is factory deactivated.
5
)
MAC address of the wired network nameplate.