The "DC Output Level" variable is used for assigning a defined offset level to the
analog output signal, e.g. for setting a certain zero offset signal level.
A "DC Output Level" >0 reduces the output range of the measurement
range 0 … 10 V by the selected offset value (F.S.R. - DC output level).
The "DC Output Level" parameter (16-Bit) consists of the high and low byte.
<data_value>×a
DC Output Level =
b
×F.S.R._Mantisse×10
(F.S.R._Exp)
Parameter
1)
Description
DC Output Level DC-Output-Signal in the selected pressure unit
( 13, table "Parameter").
1)
Further parameter 13, table "Parameter"
<data_value>
Software version =
20
e.g. <data_value> = 20 V1.0
Parameter Description
<data_value> 1 byte value (8 bit), data value in decimal format.
Parameter Description
<data_value> Zero offset measurement data, consisting of "low and high
byte" (16 bit value), data value in decimal format.
a Conversion factor for pressure units other than "Torr"
Torr: a = 1.00
mbar: a = 1.3332
Pa: a = 133.32
b Factor for resolution
Page No.
1)
b Output signal
2 8’192’000 10.20 V
3 8’192’000 10.20 V
4 8’388’352 10.00 V
F.S.R._Mantisse F.S.R. factor according to the "Sensor type" variable, which
has to be read separately ( "Read command").
F.S.R._Exp F.S.R. exponent according to "Sensor type" variable, which
has to be read separately ( "Read command").
1)
4, tabel "Structure of send string", byte no. 1
DC Output Level
Software version
Parameter