Owner manual

10
tira90e1 (2014-01) RS232C.cp
Parameter name Data
type
Access
Right
Byte
Type
Byte
No. 1
Byte
No. 2
Byte
No. 3
Comment
SystemDateTime uint8 R Byte 0 0xE2 0x00 <value> Set Date and Time.
W 0xF2
uint8 R Byte 1 0xE2 <value> Format: dd.mm.yyyy hh:mm:ss
W 0xF2 e.g. 21/10/2013 10:41:00
uint8 R 0xE2 <value>
W 0xF2
uint8 R Byte 19 0xE2 <value>
W 0xF2 (Last digit: null terminator)
EthernetLAN uint8 R 0xE2 0x14 0 Switch Ethernet LAN off
W 0xF2 1 Switch Ethernet LAN on
WirelessLAN uint8 R 0xE2 0x28 0 Switch Wireless LAN off
W 0xF2 1 Switch Wireless LAN on
WLANSettings uint8 R Byte 0 0xE2 0x51 <value> TCP-IP Address of current wifi
connection
uint8 R Byte 1 0xE2 <value> Format: IP-Adress|Network-Mask
uint8 R 0xE2 <value> Gauge resets after successful IP
change
uint8 R Byte 19 0xE2 <value> (Last digit: null terminator)
LANSettings uint8 R Byte 0 0xE2 0x65
W 0xF2
<value> TCP-IP Address of current LAN
connection
uint8 R Byte 1 0xE2
W 0xF2
<value> Format: IP-Adress|Network-Mask
uint8 R 0xE2
W 0xF2
<value> Gauge resets after successful IP
change
uint8 R Byte 19 0xE2
W 0xF2
<value> (Last digit: null terminator)
CustomAnalogOut uint8 R 0xE2 0x89 0 Switch Zoom for analog out off
W 0xF2 1 Switch Zoom for analog out on
AnalogoutPLow real32 R Byte 0 0xE2 0x81
W 0xF2
<value> Low pressure threshold for zoom
function
R Byte 1 0xE2
W 0xF2
<value> Will set this pressure value to 0.0V
R Byte 2 0xE2
W 0xF2
<value>
R Byte 3 0xE2
W 0xF2
<value>
AnalogoutPHigh real32 R Byte 0 0xE2 0x85
W 0xF2
<value> IEEE Standard for Floating-Point
Arithmetic
R Byte 1 0xE2
W 0xF2
<value> Will set this pressure value to 10.0V
R Byte 2 0xE2
W 0xF2
<value>
R Byte 3 0xE2
W 0xF2
<value>
Pressure real32 R Byte 0 0xE2 0x8B <value> Pressure value in pressure units
R Byte 1 <value> IEEE Standard for Floating-Point
Arithmetic
R Byte 2 <value>
R Byte 3 <value>
(continued)