22 Operation
6.6 Software update
6.6.1 Updating the software of the control unit
1 Copy these files to the main directory of a USB stick:
– Handset_IFC_Vx.x.xx.exe
– Handset_IFC_Vx.x.xx.key
2 Plug the USB stick into the USB port on the control unit.
3 Select on measurement screen:
F
UNCTIONS > DATA > UPDATE > CONTROL UNIT
4 Press OK to confirm message.
5 Press S
TART to begin.
6.6.2 Updating the software of the MS module
1 Copy these files to the main directory of a USB stick:
– Flash_LDS3000_MSB_Vxx.xx.xxx.bin
2 Plug the USB stick into the USB port on the control unit.
3 Select on measurement screen:
F
UNCTIONS > DATA > UPDATE > MSB
4 Press O
K to confirm message.
5 Press S
TART to begin.
6.6.3 Updating the software of the I/O module
Information The software of the I/O module can be updated from the control unit if
the mass spectrometer module has the software version “MS-Modul
1.02” or higher.
1 Copy these files to the main directory of a USB stick:
– Flash_LDS3000_IO_Vxx.xx.xxx.bin
2 Switch DIP switch S2-3 of the I/O module from OFF to ON (rising flank).
3 Plug the USB stick into the USB port on the control unit.
4 Select on measurement screen:
F
UNCTIONS > DATA > UPDATE > I/O MODULE
5 Press OK to confirm message.
6 Press S
TART to begin.
7 Switch DIP switch S2-3 of the I/O module to OFF.