Manual

20 Operation
FUNCTIONS > DATA > RECORDER > SETTINGS
FUNCTIONS > DATA > PARAMETERS
6.2.4 Access Ctrl
ACCESS CTRL > OPERATOR/SUPERVISOR/INTEGRATOR > ERROR INFO
ACCESS CTRL > OPERATOR/SUPERVISOR/INTEGRATOR
ACCESS CTRL > OPERATOR/SUPERVISOR/INTEGRATOR > PIN ASSIGN
Data recording Activation or deactivation of data recording
Off, On
Record interval Time interval between the recording of data sets
100ms, 200ms, 500ms, 1s, 2s, 5s
Memory location Memory location of the measured data that is recorded
(USB stick, control unit)
USB The data are stored on a USB stick as a TXT
file. Each TXT file contains the following
information:
- Creation date
- Software versions
- Serial number
- Start time
- Time stamp (measurement indicates offset
in seconds in relation to start time)
- File name
- Time stamp (offset in seconds in relation to
start time)
- Leakage rate (in used unit)
- Pressure P1 (in used unit)
- Device status
internal The memory in the control unit is limited
(sufficient for recording data for a maximum
of 24 hours).
Recommendation: Record directly to USB
stick.
List List of parameters that can be changed (sorted
alphabetically)
-
Error info Viewer/
Operator/Supervisor
Depth of error information
Error number, Short text, Extended information
Para. access List of parameter access levels
-
Operator/Supervisor/
Integrator
PIN for the access levels
0000 ... 9999