T R A N S L A T I O N O F T H E O P E R A T I N G M A N U A L O R I G I N A L Type designation CU1000 Product description Control unit Catalog no. 560-320 from software version Control unit 1.01 Document no.
This document applies to the software version stated on the cover page. If you need a different version, please contact our sales staff. Reprint, translation and duplication need to be approved in writing by INFICON GmbH.
Content 1 About this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 1.1 1.2 1.3 2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 2.1 2.2 2.3 3 Construction of the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.4 6.5 6.6 7 Taking out of service . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 7.1 7.2 4 Content Copying or deleting measured data . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Lock all access levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Software update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 6.6.1 Updating the software of the control unit . . . . . . . . . . . . . . . . 22 6.6.
1 About this manual 1.1 Target groups This operating manual is intended for the operator and for technically qualified personnel with experience in leak detection technology and integration of leak detection devices in leak detection systems. In addition, the installation and use of the unit require knowledge of electronic interfaces. 1.2 Other applicable documents Installation manual for mass spectrometer module Bus module installation manual I/O module installation manual Interface protocols 1.
1.3.2 6 Text markings Marking Meaning Requirement for execution of an action Tool or aid for an action ► Instruction 1, 2, 3, ...
2 Safety 2.1 Intended use The unit is intended for querying and configuring the data of the mass spectrometer module LDS3000. ► Install, operate and service the unit only in compliance with this manual. ► Comply with the limits of application (see Chapter 4.3, page 10). 2.2 User requirements Safety conscious operation ► Operate and install the unit only if it is in perfect working order and as intended, in a safety-conscious manner and fully aware of dangers, in compliance with this manual.
3 Shipment check, transport, storage 3.1 Checking shipment Scope of delivery Article Control unit Touch pin Operating manual Quantity 1 1 1 ► Check shipment to make sure it is complete. 3.2 Transport Damage due to unsuitable packaging material Transport in unsuitable packaging material can damage the unit. ► Transport the unit only in the original packaging material. ► Keep original packaging material. 3.3 Storage ► Always store the unit in compliance with the technical data, see Chapter 4.
4 Description 4.1 Construction of the unit Fig: 1 Front view 1 - Touch display Display that lets you read and enter data 2 - Status LED The Status LED indicates the status of the control unit. Status LED illuminated Status LED flashing Control unit operates normally Display is set to power saving mode 3 - Reset key Key to rest the control unit (can be operated with touch pin) Information The MS module is not reset.
Fig: 2 Rear view 1 - Rating plate Rating plate of the control unit 2Connection for headphones (impedance >= 8 Ω) or active speaker 3 - LCD-CAL Calibration key for calibrating the touch display (can be operated with touch pin) 4 - LD Connection for the data cable to the mass spectrometer module 4.2 Function You can use the control unit to configure the mass spectrometer module LDS3000. It also lets you output the data stored in the MSB box. 4.3 Technical data 4.3.1 Mechanical data 4.3.
4.3.3 Ambient conditions Permissible ambient temperature (during operation) 10 °C ... 45 °C Permissible storage temperature -20 °C ... 60 °C Max. relative humidity 4.3.4 < +31 °C 80% +31 °C to +40 °C decreasing linearly from 80% ... 50% > +40 °C 50% Type of protection IP 20 Pollution degree II Max. altitude above sea level 2000 m Factory settings The parameters of the control unit are stored in the control unit.
Description Parameter Factory setting Linear or logarithmic Log.
5 Installation 5.1 Connecting the control unit Data cable Mass spectrometer module installed. 1 Connect data cable to CONTROL UNIT of the MSB box. 2 Connect data cable to LD. Information The data cable to the control unit can also be connected or disconnected during operation. Hearing damage from loud volume setting Loud volume setting can damage hearing. ► Do no set volume of headphones too loud. 3 If needed, connect headphones or speakers to 5.2 .
6 Operation Damage to touch display as a result of incorrect operation The touch display can be damaged by a hard or pointed object. ► Operate the unit only with the fingers. 6.1 Elements on the monitor 6.1.1 Measurement display Fig: 4 1- Measurement display Keyboard lock Press and hold 2- to lock or unlock the control unit. Communication status : The unit communicates with the MS module. : The unit does not communicate with the MS module. Establish communication: 1 Reset control unit.
3The measurement is recorded. 4 - Ser The logged-in user is identified by an abbreviation. Display Ope Sup Int Ser User Operator Supervisor Integrator Service For more information, see see Chapter 6.2.4, page 20. 5 - Zero Background suppression is activated. 6Active warnings are stored in the unit. The active warnings can be displayed via the menu "INFO > HISTORY > WARNINGS". 7 - Test gas Configured test gas. Display He H2 M3 Test gas Helium (4He) Hydrogen e.g.
13 - Key FAVORITE 2 You can assign preferred parameters to this key (see Chapter 6.2.2, page 18). In Fig: 4 the key FAVORITE 2 is assigned the function START/STOP for example. 14 - Key FAVORITE 1 You can assign preferred parameters to this key (see Chapter 6.2.2, page 18). In Fig: 4 the key FAVORITE 1 is assigned the function ZERO for example. 15 - Menu All functions and parameters of the control unit can be accessed using the MENU key.
6.2 Querying and setting parameters The following chapter describes the parameters and functions of the CU1000 control unit. You will find the parameters and functions of the mass spectrometer module you can set using the control unit in the installation manual of the mass spectrometer module. The menu tree shows the access level needed for changing the parameters (factory settings).
DISPLAY > MEAS. VIEW Measurement view mode Type of graphic display Diagram, bar graph Show value Numeric representation of the measurements Off, On DISPLAY > BRIGHTNESS Display brightness 6.2.2 Brightness of the display 20 ... 100% Settings SETTINGS > TRIGGER > TRIGGER SEL. Used trigger Leakage rate limit shown in the display 1 ... 4 SETTINGS > FAVORITES Favorite 1 To execute a frequently used function more quickly, you can assign the function to the key FAVORITE 1.
SETTINGS > SET UP > CONTROL UNIT > AUDIO Volume level Volume of the headphones or active speaker 0 ... 15 Audio alarm mode Type of audio alarm No function Proportional The frequency of the audible signal is proportional to the bar graph display or diagram height. The frequency range is 300 Hz to 3300 Hz. Setpoint The pitch is proportional to the leakage rate. A signal sounds if the leakage rate exceeds the selected trigger value.
FUNCTIONS > DATA > RECORDER > SETTINGS Data recording Record interval Memory location Activation or deactivation of data recording Off, On Time interval between the recording of data sets 100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s Memory location of the measured data that is recorded (USB stick, control unit) USB The data are stored on a USB stick as a TXT file.
Information A key switch can be used instead of the PIN. The key switch is connected to the I/O module via three digital inputs. Access Ctrl Level Designation 1 Viewer (VIE) 2 Operator (OPE) 3 Supervisor (SUP) 4 Integrator (INT) 6.2.5 Login with PIN No Yes Yes Yes Parameter query Yes Yes Yes Yes set No Yes Yes Yes Info INFO > MEASUREMENT/TEMPERATURE/ENERGY/HISTORY/CONTROL UNIT/MS MODULE/ INTERFACES Diverse parameters 6.3 Information on diverse parameters, e. g.
6.6 Software update 6.6.1 Updating the software of the control unit 1 Copy these files to the main directory of a USB stick: – Handset_IFC_Vx.x.xx.exe – Handset_IFC_Vx.x.xx.key 2 Plug the USB stick into the USB port on the control unit. 3 Select on measurement screen: FUNCTIONS > DATA > UPDATE > CONTROL UNIT 4 Press OK to confirm message. 5 Press START to begin. 6.6.2 Updating the software of the MS module 1 Copy these files to the main directory of a USB stick: – Flash_LDS3000_MSB_Vxx.xx.xxx.
7 Taking out of service 7.1 Disposing of the control unit The unit can be disposed of by the user or sent to INFICON. Information The unit is made of materials that can be reused. By recycling these materials you reduce waste and environmental impact. ► For disposal, always comply with local and regional environmental and safety regulations. 7.2 Returning the control unit Danger to health Contaminated units can damage the health of INFICON's service staff. ► Use the Declaration of Contamination.
INFICON GmbH, Bonner Strasse 498, D-50968 Cologne, Germany UNITED STATES TAIWAN JAPAN KOREA SINGAPORE GERMANY FRANCE UNITED KINGDOM HONG KONG Vi sit o ur we bsi t e fo r con t act i nfo rm ati on an d o t her sal es o ffice s wo rl dw id e. w w w . i n f i c o n .