Owner manual
3 - 3
074-398-P1A
Crystal12 Sensor Operating Manual
3.3 Electrical and Pneumatic Connections
3.3.1 Electrical
To complete installation of the assembly, make electrical connections where
indicated in Figure 3-3 on page 3-5 to either 24 V(ac) or V(dc). Current required is
approximately 70 mA.
CAUTION
Maximum applied voltage must not exceed 26 V(ac)
3.3.2 Pneumatic Solenoid Tube Connections
Figure 3-1 Pneumatic Solenoid Tube Connections
TO AIR
FITTING OF
FEEDTHROUGH
EXHAUST
(Normally Open)
A OUTPUT
PORT
P SUPPLY
(Normally Closed)
80-90 PSIG
AIR SUPPLY
TUBE FITTING
(Provided with Valve)
Figure 3-1a
Solenoid Valve
Without Orifice
(As Supplied)
Figure 3-1b
Solenoid Valve
With Orifice
(Installed by User)
EXHAUST
(Normally Open)
Figure 3-1b shows the proper installation for
all CrystalSix and Crystal12 applications.
A OUTPUT
PORT
P SUPPLY
(Normally Closed)
TO AIR
FITTING OF
FEEDTHROUGH
80-90 PSIG
AIR SUPPLY
TUBE FITTING
(Provided with Valve)
ORIFICE (059-189)
(Provided with Accessory Kit 750-268)