Owner manual

2 - 9
074-398-P1A
Crystal12 Sensor Operating Manual
2.4.2 Wiring the Relay Outputs with Relay I/O Boards 760-162-G1 or
760-162-G1/G2
On the relay connector, connect the leads of the solenoid valve across the one side
of a 24 volt supply and one side of a relay (1-8, 1-16, or 1-24; whichever was
programmed). See Table 2-1.
CAUTION
Both the isolated 24 V(dc) supply and the RS-232C
remote communications port use a 9-pin D-sub
connector. Care must be taken not to inadvertently
connect the RS-232C remote communications cable to
the 24-volt supply connector. Also, care must be taken
not to inadvertently connect the 24 V(dc) supply cable to
the RS-232C remote communications connector.
Connect a jumper between the second side of the 24 volt supply and the second
side of the selected relay. See Table 2-2.
Table 2-1 +24 V(dc) Pin Connections
Pin Function
1 Return
2 Return
3 Return
4 Not Connected
5 Not Connected
6+24 Volts
7+24 Volts
8+24 Volts
9 Not Connected