User Manual
6 - 3
IPN 074-289L
Composer Operating Manual
6.2.2 Category 2; Acoustic Generator Related
NOTE: PRT stand for Platinum Resistance Thermometer, C stands for
concentration, µP stands for microphone.
6.2.3 Category 3; Acoustic Receiver Related
Table 6-2 Category 2; Acoustic Generator Related
Failure Symptom Visibility Effect on
Accuracy
Drive µP fails/not
connected
Low amplitude, Noisy C signal Noise on C
signal
Both µPs fail/not
connected
Control and Warning LEDs blink,
HALT LED illuminated
N/A
Unbalanced µP
assembly
Low amplitude or Noisy C signal Noise on C
signal
Table 6-3 Category 3; Acoustic Receiver Related
Failure Symptom Visibility Effect on
Accuracy
Receiver µP fails/not
connected
Control and Warning LEDs blink,
Error relay opens
N/A
Receiver µP loses
sensitivity
Noisy C signal at low P Noise on C
signal
Receiver µP loses bias No signal, Error relay opens N/A
Receiver µP has
unstable bias
Noisy C signal Noise on C
signal