Owner's manual
16 - 12
IPN 074-508-P1C
CMS5000 Operating Manual
27 Enter the factor of the standard in the Conc/Factor column in the Data Files
section on the left side of the screen. This number will multiply each
Concentration entered into the library template. For example, 1 µL of a
specific compound in a standard has a concentration of 1 ppb. If the standard
was prepared using 5 µL, the factor of the standard would be 5, because five
times the amount of original concentration was used. See Figure 16-24
NOTE: For more details about Concentration Reference or Conc/Factor, see
section 16.3, Glossary of Terms in the Calibrate Window.
Figure 16-24 Conc/Factor