Owner's manual
16 - 7
IPN 074-508-P1C
CMS5000 Operating Manual
12 Each individual compound will need to be manually added to the library
template. To select a peak, refer to the instructions in section 12.8, How to
Access the Scan Cursor, on page 12-25. See Figure 16-10.
Figure 16-10 Selecting a Peak
13 Once a peak is selected, right-click and select Grab Peak for Template to add
a compound to the library template. See Figure 16-11.
Figure 16-11 Grab Peak for Template
14 The following table will be displayed in the center of the screen. Enter the name
of the analyte in the highlighted field. See Figure 16-12.
NOTE: The following screenshots depict the calibration of a water method.
However, the same steps are used when creating air and check
standard methods.
Figure 16-12 Entering Analyte Name