8 Click the Browse button under Data Files. See Figure 16-6.
Figure 16-6 Browse Data Files
9 Highlight the data file of the standard with the highest concentration and click
OK. This data file will be used to create the library template. See Figure 16-7.
NOTE: The file selected below is for example only.
Figure 16-7 Selecting Data File
NOTE: The best practice for calibration library development is to build the
template using a standard with a high concentration. Low
concentration standards may have insufficient peak areas that cannot
be detected with current search settings.