Owner's manual

15 - 14
IPN 074-508-P1C
CMS5000 Operating Manual
15.6.1 Setting Up a Quantitative Search
Once a calibration library has been created (see Chapter 16, Creating Calibration
Libraries), the Library Search Parameters button will be activated. See Figure
15-19. The Library Search Parameters function sets the peak identification
criteria of the library compounds as well as the unknown analytes.
Figure 15-19 Library Search Parameters Button
Clicking the Library Search Parameters button on the Method Editor Search
page will display the window shown in Figure 15-20.
Figure 15-20 Library Search Parameters Window
The Peak Search section constitutes the parameters used in distinguishing a peak
from baseline. The four options most useful for adjusting the Peak Search are
described below.