Owner's manual
12 - 18
IPN 074-508-P1C
CMS5000 Operating Manual
12.7.5.2 Show/Update Current Peaks
The second option on the Peaks submenu is Show/Update Current Peaks. When
the base points have been edited (see section 12.7.5.3, Edit Base Points, on page
12-18), Show/Update Current Peaks will recalculate the scan range, peak
response, area and percent area utilizing the new base points.
1 Right-click in the chromatogram. Select Peaks and then, Show/Update
Current Peaks. Refer to Figure 12-27.
2 The following window will be displayed. See Figure 12-31.
Figure 12-31 Show/Update Current Peaks
12.7.5.3 Edit Base Points
The third option on the Peaks submenu is Edit Base Points. Editing the base
points will change the area of integration. Each purple dot on the chromatogram
signifies a base point. Follow the instructions below to edit base points.
1 Follow the instructions in Search for Peaks, see section 12.7.5.1 on page
12-15.
2 Right-click in the chromatogram. Select Peaks from the drop down menu and
Edit Base Points from the submenu. Refer to 12.7.5.
3 The cursor will become a plus sign. See Figure 12-32.