Owner's manual
11 - 51
IPN 074-508-P1C
CMS5000 Operating Manual
adjust the calculated the concentration. This should not be changed. The Ck Std
Multiplier Min box shows the lowest acceptable multiplier minimum value and the
CkStd Multiplier Max box shows the highest maximum multiplier value. If the
CKStd Multiplier is out of range (below the Min or above the Max) the Check
Standard method should be recalibrated, because no further adjustments to the
calibration data can be made within the acceptable limits. Refer to section 5.2,
Calibrating the Check Standard, on page 5-1.
Figure 11-87 Check Standard Autocalibration
The Run Time Prompt Enable box has two check box options. See Figure 11-88.
The Method Ready option, will pause the system before the method starts. It will
remain paused until the user selects RUN. The other option, Method Finished, will
pause the system at the end of the run. It will remain paused until the user selects
ESC to exit the screen or RUN to start a new method.