2 - 10
IPN 074-544-P1B
CI-100 Crucible Indexer Operating Manual
2.9 RS-232/LAN Connector
A DB-9 to RJ-45 adapter and standard Ethernet cable are supplied for connecting
the CI-100 to the PC for RS-232 communications.
NOTE: CI-100 does not support Ethernet communications.
2.10 Motor Connector
The controller cable supplied is a DB-25 male to female straight-through 26 AWG,
shielded. The standard length is 10 ft. (3 m). For longer lengths, use larger gauge
wire to avoid power loss to the motor drive.
Table 2-3 RS-232/LAN Connector
RJ-45 Pin Signal DB-9F Pin
1 (Wht/Org) Unused
2 (Org) Unused
3 (Wht/Grn) Unused
4 (Blu) RS-232 Tx 2 (Red)
5 (Wht/Blu) RS-232 Rx 3 (Grn)
6 (Grn) Unused
7 (Wht/Brn) RS-232 Ground 5 (Brn)
8 (Brn) Unused
Table 2-4 Motor Connector
Pin Signal
1DAC CSp
2 DAC CLKp
3DAC DATAp
4 STEPp
5DIRp
6 Encoder CLKp
7 Encoder CSp
8 Encoder DATAp
9Ground
10 Ground
11 Ground
12 24 V (dc)
13 24 V (dc)
14 DAC CSn