For reasons of compatibility, the expression "sensor cable" is used for all
BPG402 versions in this document, although the pressure reading of the
gauges with fieldbus interface (BPG402-SD or BPG402-SP) is normally
transmitted via DeviceNet or Profibus.
The sensor cable is required for supplying all BPG402 types with power.
It also permits access to the relay contacts of the switching functions
( 19, 20).
The application and length of the sensor cable have to be considered when deter-
mining the number and cross sections of the conductors ( 11).
Open the cable connector (D-Sub, 15 pins, female).
Prepare the cable and solder/crimp it to the connector as indicated in the
diagram of the gauge used:
3.2.2.1 Making an Individual
Sensor Cable
Cable type
Procedure