User guide
18 tira47e1 (2005-08) BPG402SPv1.cp
If, in the Exception Status, one of bits 4 … 6 is set, the current warning can be read
in the parameter "Exception Detail Warning". The Exception Detail Warning
parameter consists of a total of 11 bytes that inform on the error status of the
gauge.
Due to the use of the expanded error output format, these bytes have the following
structure:
Byte
No
Name Description Value
0
Common Exception
Detail Size
Indicates the number of subse-
quent bytes that contain the
Common Exception Detail
Warning.
2
1 Common Exception
Detail 0
Contains current error messages
from the Common Exception
Detail Warning group.
Depending on
warning status
2 Common Exception
Detail 1
Contains current error messages
from the Common Exception
Detail Warning group.
Depending on
warning status
3
Device Exception
Detail Size
Indicates the number of subse-
quent bytes that contain the
Device Exception Detail Warning.
5
4 Device Exception
Detail 0
One of N Status Extension. Depending on
warning status
5 Device Exception
Detail 1
(Pirani)
This error information from the
Device Exception Detail Warning
group refers to Pirani.
Depending on
warning status
6 Device Exception
Detail 2
(Pirani)
This error information from the
Device Exception Detail Warning
group refers to Pirani.
Depending on
warning status
7 Device Exception
Detail 3
(BA)
This error information from the
Device Exception Detail Warning
group refers to BA.
Depending on
warning status
8 Device Exception
Detail 4
(BA)
This error information from the
Device Exception Detail Warning
group refers to BA.
Depending on
warning status
9
Manufacturer
Exception Detail
Size
Indicates the number of subse-
quent bytes that contain the
Manufacturer Exception Detail
Warning.
1
10 Manufacturer
Exception Detail
Contains current error messages
from the Manufacturer Exception
Detail Warning group.
Depending on
warning status
3.1.1.12 Exception Detail Warning