The interface works in duplex mode. A nine byte string is sent continuously without
a request approx. every 20 ms.
Commands are transmitted to the gauge in a five byte input (receive) string.
Data rate
9600 Baud set value, no handshake
Byte
8 data bits
1 stop bit
TxD
Pin 13
RxD
Pin 14
GND
Pin 5
(Sensor cable connector)
The complete output string (frame) is nine bytes (byte 0 … 8). The data string is
seven bytes (byte 1 … 7).
Byte No. Function Value Comment
0 Length of data string 7 set value
1 Page number 5 for BCG450
2 Status
Status byte
3 Error
Error byte
4 Measurement high byte 0 … 255
Calculation of pressure value
5 Measurement low byte 0 … 255
Calculation of pressure value
6 Software version 0 … 255
Software version
7 Response value 13 for BCG450
8 Check sum 0 … 255
Synchronization
Synchronization of the master is achieved by testing three bytes:
Byte No. Function Value Comment
0 Length of data string 7 set value
1 Page number 5 for BCG450
8 Check sum of bytes 1 … 7 0 … 255 Low byte of check
sum
1)
1)
High order bytes are ignored in the check sum.