Operating Manual Instruction Manual

5 - 25
PN 074-519-P1C
3000 Micro GC Operating Manual
Figure 5-32 Installing the heated sample instroduction accessory on 3000 Micro GC
3 Tighten the 3000 Micro GC cover screws.
4 Insert the transfer line into the 3000 Micro GC sample inlet fitting as far as
possible, then back it out about 1 mm.
5 Finger tighten the female nut, then tighten an additional 1/8 turn with the
5/16 in. open-ended wrench.
6 If the transfer line is kinked, remove the sample introduction accessory and trim
about 1 mm off the end of the transfer line. Make the cut square and
perpendicular. Reassemble. See Figure 5-33.
Figure 5-33 Trimming the transfer line
7 Install and tighten the upper left cover mounting screw using T20 Torx
screwdriver.
8 Tighten the remaining cover screws with a T20 Torx screwdriver.
9 Wrap one of the two pieces of insulation around the transfer line. Retain the
remaining transfer line, insulation and extra mounting hardware as spares.
Sample inlet fitting
Transfer line
Trim 1 mm of
transfer line
Transfer line