Operating Manual Instruction Manual
5 - 17
PN 074-519-P1C
3000 Micro GC Operating Manual
Figure 5-22 shows the Pressure Reducer and Genie Filter Assembly mounted
on 3000 Micro GC.
Figure 5-22 Pressure reducer and Genie filter assembly mounted on 3000 Micro GC
5.1.6 Installing Heated Sample Introduction Accessories
Heated sample introduction accessories are separately powered sample
conditioners for the 3000 Micro GC. See Table 5-5.
G2816A on a 2-channel 3000 Micro GC
Table 5-5 Heated sample introduction accessory types
Model PN Description Source Pressure
G2818A-X/G2845
A-X/G2857A-X
Heated Regulator for natural gas at
variable pressure
14 to 5516 kPa (2 to 800 PSI)
G2819A-X/G2846
A-X/G2858A-X
Heated Vaporizer for liquefied
petroleum gas under high pressure
1379 to 5516 kPa (200 to 800 PSI)
NOTE: G2818A-X and G2819A-X are heated sample introduction accessories designed for a
1,2-channel 3000 Micro GC.
NOTE: G2845A-X and G2857A-X are heated sample introduction accessories designed for a
3,4-channel 3000 Micro GC.
NOTE: G2857A-X and G2858A-X are heated sample introduction accessories designed for a
Portable 3000 Micro GC.
NOTE: The Heated Regulator and Heated Vaporizer come equipped with a separate power
cord. The PN reflects the power cord option with an X, where X is a number representing
one of the following countries:
1: China
2: Europe
3: US
4: Japan
5: UK/HK/SG/MY
6: Australia/NZ
7: Korea