Operating Manual Instruction Manual

5 - 13
PN 074-519-P1C
3000 Micro GC Operating Manual
5.1.5.3 Installation Instructions for the Pressure Reducer and Genie Filter Assembly
1 Turn off 3000 Micro GC and wait for the sample inlet fitting to cool.
3000 Micro GC has inlets which can be heated to 140°C.
Touching the inlets at operating temperatures can result
in injury. Extreme care should be taken to avoid touching
the heated inlets.
2 Move the 3000 Micro GC handle fully away from the front cover.
3 Remove any fittings or hardware currently installed on the 3000 Micro GC
sample fitting.
4 Place a 1/4 in. male nut, back ferrule, and front ferrule onto one end of the outlet
tube provided. See Figure 5-16 for ferrule orientation. See Figure 5-17 for outlet
tube orientation.
5 Loosely install a 5/16 in. nut, back ferrule, and front ferrule onto the sample inlet
fitting. Do not tighten. The outlet tube will be installed into this fitting later.
Figure 5-16 Installing ferrules
1/4 in. male nut
Back ferrule
Front ferrule
Back ferrule
Front ferrule
5/16 in. female nut