Operating Manual Instruction Manual
5 - 1
PN 074-519-P1C
3000 Micro GC Operating Manual
Chapter 5
Accessory Installation
5.1 Sample Conditioners
3000 Micro GC can be damaged by contaminants, especially particulates and
condensing aerosols. Use an appropriate filter or sample conditioner at all times.
Table 5-1 lists the typical usage and filtration capabilities of the standard filter and
available accessories.
5.1.1 List of Sample Conditioners and IPNs
Table 5-1 List of sample conditioners, inlet pressures, and IPNs
Type Input Pressure
to Accessory
Sample
Container/delivery
Sample Matrix Particle
filtration
(microns)
PN
Standard external
sample inlet filter, or
direct connection
0 to 172 kPa
(0 to 25 PSI)
All Relatively clean
and dry
10 G2801-60900
Genie Filter
Assembly
70 to 345 kPa
(10 to 50 PSI)
All Entrained liquids
and particles
- G2817A
Pressure Reducer 345 to 6895 kPa
(Ambient to
1000 PSI)
All C6+ components
<0.5 mole%
- G2815A
Pressure Reducer
and Genie Filter
Assembly
<3477 kPa
(<500 PSI)
All Entrained liquids
and particles,
C6+ components
<0.5 mole%
G2816A