Operating Manual Instruction Manual
9 - 6
PN 074-519-P1C
3000 Micro GC Operating Manual
TCD and columns turn off
(continued)
High carrier gas pressure Ensure the carrier gas is supplying at least
78 PSI, and no more that 82 PSI.
CAUTION
Exceeding 82 PSI for the
carrier gas pressure may
result in damage to
3000 Micro GC.
Malfunctioning column or
injector
Check the flow coming out of the analytical
and reference vents on the back of the
instrument. If there is no flow on either/both
vents, contact INFICON.
Malfunctioning TCD Contact INFICON.
Malfunctioning EPC valve Contact INFICON.
Inlet/column temperature not
stabilizing
Instrument requires power
cycle
Restart 3000 Micro GC as instructed in
section 6.3.2, Restart, on page 6-8.
Hardware problem Contact INFICON.
Unable to download
method/error message
Instrument requires power
cycle
Restart 3000 Micro GC as instructed in
section 6.3.2, Restart, on page 6-8.
Incorrect configuration See section 7.2, Connecting EZ IQ with
3000 Micro GC, on page 7-1.
EZ IQ requires restart Close EZ IQ and re-open. Restart the
computer if necessary.
Method is corrupt Create a new method and reprogram
parameters.
Drivers are corrupt Configure the instrument through EZ IQ
Config.
D-Board error Contact INFICON.
TCD detector 'fault' Instrument requires power
cycle
Restart 3000 Micro GC as instructed in
section 6.3.2, Restart, on page 6-8.
EZ IQ requires restart Close EZ IQ and re-open. Restart the
computer if necessary.
Drivers are corrupt Configure the instrument through EZ IQ
Config.
Table 9-2 Common software problems (continued)
Symptom Cause Remedy