Operating Manual Instruction Manual
8 - 23
PN 074-519-P1C
3000 Micro GC Operating Manual
8.3 Sample Conditioners
8.3.1 Sample Conditioner Connections
There are multiple types of optional sample conditioners. A sample conditioner
connects directly to the 3000 Micro GC without a basic filter (G2801-60900).
Make sure the proper fittings are available.
8.3.1.1 Connections for Sample Inlet Filter (PN G2801-60900), Pressure Reducer
(PN G2815A), Pressure Reducer and Genie Filter Assembly (PN G2816A), and
Genie Filter Assembly (PN G2817A)
8.3.1.2 Connections for the Heated Regulator (PN G2818A-X/G2845A-X/G2857A-X)
The Heated Regulator requires a Swagelok quick disconnect fitting. Use the quick
disconnect fitting provided with the accessory. This quick disconnect fitting accepts
a 7/16 in. threaded male connector.
8.3.1.3 Connections for the Heated Vaporizer (PN G2819A-X/G2846A-X/G2858A-X)
The Heated Vaporizer requires the Swagelok quick disconnect fitting (included with
the Heated Vaporizer). This quick disconnect fitting accepts a 7/16 in. threaded
male connector.
Table 8-2 Input fitting for the sample inlet filter and sample conditioners
3000 Micro GC or Accessory User Fitting
GC Input Fitting or Sample Inlet Filter Assembly 1/16 in. Swagelok
G2815A Pressure Reducer 1/8 in. Swagelok
G2816A Pressure Reducer and Genie Filter Assembly 1/8 in. NTP (male)
G2817A Genie Filter Assembly 1/16 in. stainless steel tubing