Operating Manual Instruction Manual
8 - 20
PN 074-519-P1C
3000 Micro GC Operating Manual
8.2.7 Commission a New Standard GC Module
For Standard GC modules, a new module IP address must be assigned to the
newly installed GC module. This can be accomplished using the embedded
3000 Micro GC web page.
NOTE: For a Performance Enhanced GC module, these steps can be skipped.
Proceed to section 8.2.8, Enable the Instrument in EZ IQ, on page 8-22.
1 Turn 3000 Micro GC on
2 Wait three minutes for the instrument to fully power up.
3 In the web browser address bar, type the 3000 Micro GC IP address.
4 Click the Utilities tab.
5 Click Change Module Configuration.
Figure 8-24 Change module configuration
6 Click Set. See Figure 8-25.
Figure 8-25 Commission a standard GC module - set