3000 Micro GC Operating Manual
8.2.6 Install a GC Module
Once a module is removed, it can be replaced with another module. If the
replacement module uses a different carrier gas, follow the procedure in section
6.4, Change Carrier Gas Configuration, on page 6-9.
NOTE: For the Standard GC Module, skip step 1 and proceed to step 2.
1 Set the Performance Enhanced GC module address by setting the DIP switch
on the upper right of the D-board to the appropriate configuration. See Figure
8-17.
Figure 8-17 DIP switch position on the D-board
Each setting determines a unique module address. Refer to the label on the top
of the module. DEF is the default DIP switch setting that is set in the factory.
See Figure 8-18.
Figure 8-18 Module address label on top of a performance enhanced module