Operating Manual Instruction Manual

7 - 40
PN 074-519-P1C
3000 Micro GC Operating Manual
7.12.4 Perform a Multiple Level Calibration
If a multiple level calibration is required, sample calibration standards of different
concentration levels should be run.
NOTE: Update the Method >> Peaks/Groups table to include Level 2, Level 3,
etc.
1 Repeat Steps 1 through 7 in section 7.12.3 on page 7-37.
2 Select the Calibrate check box and type a value in the Calibration level box
for the corresponding level. See Figure 7-52.
Figure 7-52 Analysis/single level calibration window - multiple level calibration
3 Click Start.