Operating Manual Instruction Manual

7 - 14
PN 074-519-P1C
3000 Micro GC Operating Manual
4 The Instrument Setup window will display parameters that can be changed.
See Figure 7-17.
NOTE: The Instrument Setup window is configuration dependent.
Figure 7-17 Instrument setup window
5 Enter sample acquisition parameters as desired, such as Injector
Temperature, Column Temperature, Sample Pump time, Column
Pressure, etc. Ensure that the Run Time is long enough for the last expected
peak to elute. If the elution time is unknown, conduct a run and verify the
presence of the last expected peak. If necessary, increase the Run Time.
The default Inject Time should be set between:
0-50 ms for a fixed volume injector
0-100 ms for a variable volume injector
0-400 ms for a variable large volume injector (LVI)
0-250 ms for a backflush injector