Operating Manual Instruction Manual

6 - 12
PN 074-519-P1C
3000 Micro GC Operating Manual
11 Click Continue from the 3000 Micro GC status view to implement the
configuration change.
12 Change the gas supply connections at the 3000 Micro GC back panel, and
check for leaks.
13 Click Restart. See Figure 6-16. Wait at least three minutes before shutting
down the instrument.
Do not turn off the power without waiting at least three
minutes. 3000 Micro GC must update configuration files
before power is turned off. If power is turned off too soon,
the files could be corrupted, making 3000 Micro GC
inoperable.
Figure 6-16 Restart screen for gas configuration
14 Turn on 3000 Micro GC and wait three minutes for it to fully boot up.
NOTE: To run samples using the reconfigured gas type, open EZ IQ Config and
update the 3000 Micro GC configuration in EZ IQ. Refer to the 074-537 EZ
IQ Installation Guide for details.
NOTE: A new EZ IQ method must now be created for the instrument.
NOTE: Changing the gas type turns off the detector filaments. Turn on the
filament(s) before use.