IDS 440 Weigh Scale Indicator Installation/Calibration/Operation Version 4.3 06/13/08 Program S440012x Industrial Data Systems, Inc. 3822 E. La Palma Ave. Anaheim Ca. 92807 Tech Support 714-921-1353 Website: www.industrialdata.com Email: Sales@industrialdata.
Table of Contents Introduction ......................................................................................................... 1 What’s In Each Section ..............................................................................................................2 WARRANTY INFORMATION ....................................................................................................3 General Description............................................................................................
Set Operation Parameters................................................................................ 21 Access Operation Parameters and Parameter Functions .......................................................21 Operation Parameters Reference Tables ................................................................................21 Scale Parameters.....................................................................................................................25 Parameter 1.
Function 64 Configure Passwords ..........................................................................................38 Function 65 Configure Print Formats ......................................................................................39 Function 66 Set Time and Date ..............................................................................................43 Function 67 Display Operation Parameters ............................................................................
IDS440 Users Manual version 4.2 Introduction Introduction Congratulations on your purchase and welcome to the IDS440 User's Manual. This manual describes the installation, calibration and setup, and operation of the IDS440 weigh scale indicator. The IDS440 is a microprocessor based weight indicator with rugged design and state-of-the-art technology.
IDS440 Users Manual version 4.2 Introduction What’s In Each Section This manual is organized and divided into separate sections, which describe the installation, calibration, setup, operation, and testing of the IDS440. Some Sections, such as Installation, Setup and Calibration are intended for trained scale technicians familiar with the installation of computer equipment. Other sections present information applicable to a user, installer, or system integrator.
IDS440 Users Manual version 4.2 Introduction WARRANTY INFORMATION INDUSTRIAL DATA SYSTEMS, INC. LIMITED WARRANTY. Industrial Data Systems, Inc., warrants that the products furnished are free from defects in material and workmanship and perform to applicable, published Industrial Data Systems, Inc., specifications for one (1) year from the date of shipment. THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED. IN NO EVENT SHALL INDUSTRIAL DATA SYSTEMS, INC.
IDS440 Users Manual version 4.2 General Description General Description This section gives you a general description on the operation of the IDS440's keyboard and display. The keyboard is used to initiate weighing functions, for data entry, maintenance, setup, and testing. The numeric display is used for weight and error display and for data entry. The LED status lights indicate the status of the weigh scale. Additional information about the keyboard and display is provided in section: Using the IDS440.
IDS440 Users Manual version 4.2 General Description The IDS440 Display The IDS440 DISPLAY consists of a 6 character NUMERIC display and 6 statuses LED’s The NUMERIC display is normally used to display the weight on the scale. The LED’s display the scale status. When lighted they indicate the following conditions: GROSS- The number displayed is the gross weight on the scale. NET - The number displayed is the net weight on the scale. MOTION- The weight on the scale is changing (not stable).
IDS440 Users Manual version 4.2 Installation and Setup Installation and Setup This section provides information about unpacking, installing, and setup of the IDS440. It also directs the installer to the appropriate sections of the manual for hardware and setup installation. Unpacking the IDS440 Installation begins with unpacking the IDS440. Observe any instructions or cautions, which appear in the shipping container.
IDS440 Users Manual version 4.2 Installation and Setup Connect IDS440 to host device (optional) The Serial Communications Port 2 (TB2) provides a continuous output of scale data. This is used for connection to a host device such as a computer or to a remote display (scoreboard). Continuous output transmission can be enabled or disabled and the output format can be customized to interface with the host device. See Hardware and Wiring Diagrams for pin assignments and sample cable drawings.
IDS440 Users Manual version 4.2 Installation and Setup Setup Guide Setup the IDS440 in the following order: Initialize to Factory Defaults The IDS440 is initialized at the factory. It must be re-initialized if the program memory (EPROM) or the parameter memory (EAROM) is replaced. The Initialize function clears the memory and sets the configuration parameters to their default values. Use parameter function 70 described in Parameter Functions to initialize the IDS440.
IDS440 Users Manual version 4.2 Built In Application Programs Using the IDS440 The IDS440 has 7 function keys located at the bottom of the keyboard, a numeric keypad, and 2 control keys (ENTER, and CLEAR). This section of the IDS440 manual describes what the keys are used for and how to use them. Keyboard Layout Operation of Function Keys The function keys are located on the bottom of the IDS440’s front panel. Zero Press the ZERO key to zero the scale.
IDS440 Users Manual version 4.2 Built In Application Programs Tare AUTO TARE: Press the TARE key. The IDS440 reads the weight on the scale and stores it in the tare register. The IDS440 switches to NET mode and displays the net weight. KEYBOARD TARE: Use the numeric keys to enter a tare weight. Press the TARE key. The IDS440 stores the entered weight in the tare register and switches to NET display mode. CLEAR TARE: Press the numeric ZERO key. Press the TARE key.
IDS440 Users Manual version 4.
IDS440 Users Manual version 4.2 Built In Application Programs Built In Application Programs The following built in programs are available: 1. Fill to Setpoint. 2. Checkweigh: Under/Between/Over. 3. Weigh-in / Weigh-out. 4. Axle Weigh – Short Scale. 5. Axle Weigh – Long Scale. 6. ID Tare and Total. 7. Peak Hold. Use parameter function 73, “Initialize Function Memory”, to activate a built in program.
IDS440 Users Manual version 4.2 Built In Application Programs #1 Fill to Setpoint Press F1 key to enter a setpoint. Place an empty container on the scale. Press F2 key to begin filling. TTL output 1 turns on until the weight on the scale is equal or greater than the setpoint. Installation Connect TTL output 1 to relay or lamp actuator.
IDS440 Users Manual version 4.2 Built In Application Programs # 2 Checkweigh: Under/Between/Over The indicator monitors the scale for activity. When the scale goes above the ‘empty’ weight, the indicator waits for stable weight, and then outputs an indication of Under, Between (OK), or over. The cycle resets when the scale goes below the empty weight. Installation Connect TTL outputs to relay or lamp actuators.
IDS440 Users Manual version 4.2 Built In Application Programs # 3 Weigh-in / Weigh-out This program uses the ID memory for weigh-in / weigh-out transactions. An ID record is created using the Weigh-in function. The Weigh-out function recalls the ID, calculates Gross/Tare/Net of the transaction, prints the transaction, and adds the net weight to the ID’s totals register. The ID data is saved for further weigh-in/weigh-out transactions, or weigh-out only transactions.
IDS440 Users Manual version 4.2 Built In Application Programs REBUILD ID MEMORY: Press the ENTER key when the display prompts, “rebuiLd”. The display then prompts “id”. Enter the numeric 0 then press ENTER to rebuild ID memory. COUNT UNUSED ID’S: Press the ENTER key when the display prompts “Count”. The display prompts “FrExxx” where xxx is the number of unused ID records in memory. Set Beginning Sequence Number Use the numeric keypad to enter number 9.
IDS440 Users Manual version 4.2 Built In Application Programs # 4 Axle Weigh – Short Scale Automatically weigh multiple axles (or containers), print each axle weight, and print axle total. Installation Connect the Red light / Green light relays. Green light = TTL out 1 TB4 pin 1 Red light = TTL out 2 TB4 pin 2 Setup Parameters Configure the following parameters: Parameter 43 Trip level weight. Determines when an axle is on the scale. Parameter 74, 1 Red light turn delay timer.
IDS440 Users Manual version 4.2 Built In Application Programs # 5 Axle Weigh – Long Scale, Unattended Automatically weigh multiple axles (or containers), print each axle weight, and print axle total. Installation Connect the Red light / Green light relays. Green light = TTL out 1 TB4 pin 1 Red light = TTL out 2 TB4 pin 2 Setup Parameters Configure the following parameters: Parameter 43 Trip level weight. Determines when an axle is on the scale. Parameter 74, 1 Red light turn delay timer.
IDS440 Users Manual version 4.2 Built In Application Programs # 6 ID, Tare, and Total This program uses the ID memory to store tare weights and totals. Enter ID Tare Data Press the F1 key. The display prompts "id". Enter up to 8 digits for ID and then press the ENTER key. The display prompts “tArE”. Enter the tare weight. Recall ID and Tare Press the F2 key. The display prompts "id". Enter up to 8 digits for the weigh-out ID and then press the ENTER key.
IDS440 Users Manual version 4.2 Built In Application Programs enter a new value and then press the ENTER key. The next sequence number printed will be the entered value +1. Notes • • The ID file maintenance functions are password level 1 protected. If a password has been configured (function 64) the operator will be prompted for the password before the file maintenance menu is displayed. ID memory is cleared to zero at power up unless parameter 41 (battery back up enable) is set to 1.
IDS440 Users Manual version 4.2 Set Operation Parameters Set Operation Parameters This section describes the operation parameters of the IDS440 and how their settings affect the operation of the weight indicator. Access Operation Parameters and Parameter Functions The parameters are accessed by holding the CLEAR key down and then pressing the ENTER key. The numeric display will prompt “CFG xx” where xx is the currently selected parameter.
IDS440 Users Manual version 4.
IDS440 Users Manual version 4.
IDS440 Users Manual version 4.2 Set Operation Parameters Parameter Functions Fn. No 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 78 79 80 81 82 83 Function Name Display Calibration Audit Number Calibrate Scale - Deadload first Calibrate Scale - span first Calibrate Deadload Only Adjust Gain Calibration Configure Passwords Configure Print Formats Set Time and Date Display Operation Parameters Print Operation Parameters Diagnostic Tests Initialize Operation Parameters to factory defaults.
IDS440 Users Manual version 4.2 Set Operation Parameters Scale Parameters The scale parameters configure IDS440 for the weigh platform or load cell that it is connected to. The preceding table titled “SCALE PARAMETERS” lists the scale parameters and their factory settings. Use the “Field Setting” column to record any changes made to the factory default values. Many of the following parameters are entered in terms of “graduations”.
IDS440 Users Manual version 4.2 Set Operation Parameters Parameter 4. Motion Detection Delay (6) Motion detection delay is used by the motion detection function to determine if the scale is in motion (not stable). Enter the amount of time (in tenths of a second) that the scale must be stable for motion status to be false (scale status = stable). Example: 0 = disables motion detection. 2 = 0.2 second delay before motion = false (stable = true). 15 = 1.5 second delay. 30 = 3.0 second delay. Parameter 5.
IDS440 Users Manual version 4.2 Set Operation Parameters Parameter 10. Primary Units Type (1 [lb]) The scale is calibrated in the primary units. The UNITS key is used to toggle the weight display between primary units and alternate units.
IDS440 Users Manual version 4.2 Set Operation Parameters Parameter 15. Alternate Count-by (5) Enter the alternate units count-by. Parameter 16. Alt Units Conversion Factor (45,360) The IDS440 calculates the alternate weight by multiplying the primary weight by the conversion factor and then dividing by 100,000. The following table lists some common conversions. Alternate Units Conversion Factor Table Conversion Conversion Factor Conv. Fact. x 100,000 lb to kg 0.45359 kg/lb 45,359 kg to lb 2.
IDS440 Users Manual version 4.2 Set Operation Parameters Parameter 21. Deadload Factor The deadload factor is automatically calculated by the calibration function. This parameter is not accessible from the keyboard. Parameter 22. Weight Conversion Factor The weight conversion factor is automatically calculated by the calibration function. This parameter is not accessible from the keyboard.
IDS440 Users Manual version 4.2 Set Operation Parameters Parameter 24 Serial Port 1 Baud Rate (1 [9600 baud]) Enter a baud number from the table below to select a baud rate from the description column. (*HS baud rates for the 100 samples per sec version only). Baud Number 1 2 3 4 5 6 7 Baud Description 9600 Baud 4800 Baud 2400 Baud 1200 Baud 600 Baud 300 Baud (* HS 19.2K) 150 Baud (*HS 38.
IDS440 Users Manual version 4.2 Set Operation Parameters Enter a station ID (1 to 254) for multi-drop network configuration • Enter 255 for single station remote control • Parameter 28 TX2 Format (2 Condec format) Serial Port 2 can be used for continuous transmission of scale data. The TX2 Format parameter determines what is sent.
IDS440 Users Manual version 4.2 Set Operation Parameters Printer Parameters The IDS440 provides a universal printer interface for various printers. The printer interface allows you to select and customize the data output for serial or parallel printers. The serial interface option uses Serial Port 1 (TB1) for connection to the printer. The parallel interface uses the TTL I/O Port (TB4) for connection to the printer.
IDS440 Users Manual version 4.2 Set Operation Parameters Parameters 35, 36, 37, and 38: Print Codes The IDS440 provides 4 programmable print codes. The print codes are used to send special setup parameters to the printer, such as select print size or select print font. Each print code is up to 4 characters long. Consult your printer manual for the print codes that you may want to use. Enter the codes in decimal format (an ASCII carriage return = 13). Select parameter 35 to begin print code 1 entry.
IDS440 Users Manual version 4.2 Set Operation Parameters Parameters 51-58 Keyboard Event Functions The keyboard event functions determine what actions take place when a key is pressed. The Power on start function activates a function when power is applied to the indicator.
IDS440 Users Manual version 4.2 Parameter Functions Parameter Functions The parameter functions are used to set parameters that are too complex for a single numeric entry. The following table lists the parameter functions: Parameter Functions Fn.
IDS440 Users Manual version 4.2 Parameter Functions To exit configuration and store changes, press the CLEAR key when “CFG xx” is being displayed. To exit configuration and NOT store changes, enter 999 when “CFG xx” is being displayed. The display prompts “Abort”. Press the ENTER key to abort (exit without saving changes) or press the CLEAR key if you do not want to abort. Example: Begin Calibration of the ScaleThe IDS440 is in the display weight mode.
IDS440 Users Manual version 4.2 Parameter Functions Hold the CLEAR key down and press the ENTER key to enter the configure mode. 1. Enter CFG 60. The IDS440 prompts “dEAdld”. Remove all weight from the scale. Press the ENTER key. The IDS440 reads the scale’s deadload weight. NOTE: press the CONTINUE key (Units) to skip deadload read. 2. The IDS440 display alternates between “CALib” and the weight on the scale. Place the calibration weight on the scale.
IDS440 Users Manual version 4.2 Parameter Functions 3. Press the ‘1’ key to adjust by 1, press the ‘2’ key to adjust by 10, the ‘3’ key to adjust by 100, or the 4 key to adjust by 1000. Use the UNITS key (↓) and the PRINT key (↑) to increase or decrease the displayed weight until the displayed weight is the same as the weight on the scale. The center of digit indicator (-) should be displayed for maximum accuracy. 4. Press the CLEAR key to exit Adjust Gain procedure.
IDS440 Users Manual version 4.2 Parameter Functions Function 65 Configure Print Formats Use function 65 to modify print formats to meet your print requirements. The print formats are grouped into PAGES that are used for generating print data onto a form. Each page is organized into rows and columns as follows: Line No. Column No. Item No. Line No. = Line number on the page to print the item on. Column No. = Column position where item print begins. Item No. = Item to print.
IDS440 Users Manual version 4.2 Parameter Functions Page List FORMAT PAGE 1 PAGE 2 PAGE 3 PAGE 4 DESCRIPTION Prints when the IDS440 is in the GROSS weight mode. Prints when the IDS440 is in the NET weight mode. Used by Scale Basic functions when needed. Used by Scale Basic functions when needed. Print Item List Item # 2-4 84-90 5 6 7 8 9 10 11 12 15 16 17 18 21 22 23 24 25 26 27 32 33 34 35 40 41 42-46 51-66 84-90 Description Print Label 1, 2, and 3 (see chap.
IDS440 Users Manual version 4.2 Parameter Functions Print Format Design Example EXAMPLE PRINT: COLUMN LINE 1 2 3 4 5 6 7 1111111111222222222233333333334 1234567890123456789012345678901234567890 GROSS TARE NET 54785 LB 12451 LB 42243 LB TIME 03:45 PM DATE 05 AUG 1999 EXAMPLE FORMAT DESIGN WORKSHEET: Page:1 Item Description Line No. E1 Gross Weight 3 E2 Tare Weight 4 E3 Net Weight 5 E4 Time and Date 7 E5 End of Print 0 Column 4 4 4 1 0 Item No.
IDS440 Users Manual version 4.2 Parameter Functions Format Design Worksheets Page Item Description Line No. Column Item No. Line No. Column Item No.
IDS440 Users Manual version 4.2 Parameter Functions Function 66 Set Time and Date Use function 66 to set the time and date. Entering 19 can display the time and date when the meter is in the idle mode. Date will be printed in a Y2K compliant four-digit year. 1. Hold the CLEAR key down and press the ENTER key to enter the configure mode. 2. Enter CFG 66. The IDS440 prompts “t”. Enter the time using 5 digits. The last digit should be 0 for AM, 1 for PM, or 2 for 24-hour time. 3. The IDS440 prompts “D”.
IDS440 Users Manual version 4.2 Test # 1 2 3 4 5 6 7 8 9 10 Parameter Functions Hardware Tested Serial Com. Port 1 Serial Com. Port 1 Serial Com. Port 1 Serial Com. Port 2 Serial Com. Port 2 Serial Com. Port 2 Parallel Port A/D converter Ram Memory Display Description Display input data. Display input error count. Transmit data. Display input data. Display input error count. Transmit data. Transmit data. Display raw conversion data. Test memory for errors. Turn on all LED’s and display segments.
IDS440 Users Manual version 4.2 Parameter Functions 4. Use the UNITS key (↓) or the PRINT key (↑) to scroll through the parameter numbers. Press the CLEAR key to exit from the current Event number. Press the CLEAR key again to exit from event monitor data entry. Function 72 Scale Basic Function Editor Use function 72 to create or modify Scale Basic functions. See the “Scale Basic / Event Monitor” manual for more information on Scale Basic functions. 1.
IDS440 Users Manual version 4.2 Parameter Functions Parameter 76 Debug mode for Event monitors and Scale Basic The Debug mode is used to ‘debug’ event monitors and scale basic functions. Enter a 1 to activate debug mode, enter a 0 to disable debug mode. When debug mode is enabled Scale Basic functions ‘single step’. The subroutine stack level and the Function number are displayed when entering a function. Example: “0 F12”. Press the ENTER key to begin execution of the function.
IDS440 Users Manual version 4.2 • • Parameter Functions Parameter 79 must be set to 1 to enable the IDS440 to use the multi-point Linearization calibration parameters. Use the UNITS key (↓) to continue to the next calibration point if the current point has already been calibrated.
IDS440 Users Manual version 4.2 Diagnostic Tests Diagnostic Tests The DIAGNOSTIC TESTS are accessed using configuration function 69. Hold the CLEAR key down and press the ENTER key. The display prompts “CFG xx” where xx is the currently selected parameter. Enter 69 and press the ENTER key to access the DIAGNOSTIC TESTS. The display prompts “diA xx” where xx is the last selected diagnostic function. Select a test number from the table below. Enter the test number and press the ENTER key.
IDS440 Users Manual version 4.2 Diagnostic Tests Diagnostic Test 3: Serial Com Port 1 - Transmit Data This test transmits data out of serial communications port 1. The display prompts: “OUT 0”. Press the ENTER key to send data. Serial Port 1 transmits “1234567890ABCDEF” and prompts “OUT 1” to indicate that 1 transmission has occurred. Press the ENTER key again or press the CLEAR key to exit diagnostic test 3.
IDS440 Users Manual version 4.2 Diagnostic Tests display. The data will not transmit if the BUSY signal (TB4 pin 14) is active (TTL high). Check the cable connections and the sending device. Diagnostic Test 8: Test A/D This test displays the (filtered) raw data being read from the A/D converter. Use this test to determine optimum deadload and gain settings.
IDS440 Users Manual version 4.2 Troubleshooting Troubleshooting The following table describes the probable causes to some problems you may encounter. Most problems can be resolved by using the information provided in the table. Problems And Probable Causes Problem Printer does not print (serial RS232) Continuous output does not update Input port 1 not working Input port 2 not working “Er CtS” error message.
IDS440 Users Manual version 4.2 Troubleshooting Error Messages Error Err 0 Err 1 Err 2 Err 3 Err 4 Err 5 Err 5.1 Err 6 Err 7 Err 8 Err 9 Err 10 Err 11 Err 12 Err 13 Err 14 Err 15 Err 16 Err 18 Err 21 Err 22 Err 23 Error OL undEr Description Power on zero error. See parameter 20. Keyboard error. Occurs during power up while a key on the keyboard is being pressed or one of the remote switch inputs is activated. Restart Trap. The microprocessor accessed nonexistent memory location.
IDS440 Users Manual version 4.2 Hardware and Wiring This section describes installation and wiring information for the interface ports. There are two bi-directional serial ports, one parallel port with TTL inputs and outputs, and one load cell input port. The serial ports are used to interface to a printer and to a host device or PC for continuous output. The parallel port is used to interface to parallel printers and for remote switch input and relay control using the TTL I/O.
IDS440 Users Manual version 4.2 Serial Port 1 Connector - TB 1 For RS232 input, set switch S1 to RS232 position, and for current loop input, set the switch to CL position. NOTE: Jumper pin 7 CTS to 8 RTS if unit locks up on “Print”.
IDS440 Users Manual version 4.2 TTL I/O Port Connector - TB 4 TTL outputs 1 thru 8 will drive 24ma at TTL low level, and 8ma at TTL hi level. Output 9 is an open collector output with 4.7k-ohm resistor to +5V.
IDS440 Users Manual version 4.2 P.C.
IDS440 Users Manual version 4.
IDS440 Users Manual version 4.2 DT210/IDS550/IDS710/PC (Continuous Output Port) Serial Com Port 2 TB2 2 5 DT210 / IDS550 / IDS710 (RS232) TXD GND RXD GND 3 7 DB 25 PIN MALE Serial Com Port 2 (20ma Current Loop) TB1 4 6 DT210 / IDS550 / IDS710 +5R CL1 TX CL+ CL- 8 22 DB 25 PIN MALE Hewlett Packard Smart Wand HBCR-8xxx series Serial Com Port 1 TB1 3 5 1 (RS232) Bar Code Reader HBCR-8xxx RXD GND +5V TXD GND Power 2 7 9 NOTE: set SW1 to RS232 position.
IDS440 Users Manual version 4.
IDS440 Users Manual version 4.2 IDS440 Specifications Display Keyboard 1 inch x 6-digit numeric, 6 LED status indicators. Sealed keyboard with tactile feel. I/O Ports Load Cell Interface Port. Serial port 1 - RS232 or Cur. Loop. Serial port 2 - RS232 or RS485. 9 TTL outputs. 8 TTL inputs. Display Graduations 120,000 graduations (industrial) 10,000 hb44. Display Increments 1, x2, x5, x10, x20, x50, x100. Decimal Point 0 to 4 decimal places. Internal Graduations 1,024,000.
IDS440 Users Manual version 4.2 ASCII CHART ASCII Null SOH STX ETX EOT ENQ ACK Bell BS HT LF VT FF CR SO SI DLE XON TAPE XOFF DC4 NAK SYN ETB CAN EM SUB ESC FS GS RS US DEC 00 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 ASCII SP ! " # $ % & ' ( ) * + , .