NOTE UT is the a.c. mains voltage prior to application of the test level.
Manufacturer’s declaration-electromagnetic immunity
The Yushan V14C, Yushan V14G is intended for use in the electromagnetic environment (for
professional healthcare) specified below.
The customer or the user of the Yushan V14C, Yushan V14G should assure that it is used in
such and environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment-guidance (for
professional healthcare environment)
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms:
0,15 MHz – 80
MHz
6 Vrms:
in ISM bands
between
0,15 MHz and
80 MHz
80 % AM at 1
kHz
3 V/m
80 MHz – 2,7
GHz
80 % AM at 1
kHz
3 Vrms:
0,15 MHz – 80
MHz
6 Vrms:
in ISM bands
between
0,15 MHz and
80 MHz
80 % AM at 1
kHz e)
3 V/m
80 MHz – 2,7
GHz
80 % AM at 1
kHz
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Yushan V14C, Yushan V14G including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance:
d = 1,2
√
P
d = 1,2
√
P 80MHz to 800 MHz
d = 2,3
√
P 800MHz to 2,7 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in metres (m).
Interference may occur in the vicinity of
equipment marked with the following
symbol: