B.2 Further Information for IEC
60601-1-
2 (EN 60601
_____________________________________________________________
Caution
This equipment is not intended for use in residential environments and may not
provide adequate protection
_____________________________________________________________
1. Medical electrical equipment needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information
p
rovided in the accompanying documents.
2. Portable and mobile RF communications equipment can affect medical
electrical equipment.
3. Information regarding the cable affecting EMC is as follows.
Name
Connection
Network
cable
Extend Back
cable to PC or
network switch.
Power cord
From adaptor to AC
outlet
4. The use of accessories, transducers and cables other than those specified,
with th
e exception of transducers and cables sold by
Corp.
as replacement parts for internal components, may result in increased
emissions or decreased immunity of the Yushan.
5. The Yushan shoul
d not be used adjacent to or stacked with o
equipment.
If adjacent or stacked use is necessary, the Yushan FPD should be observed
to verify normal operation in the configuration in which it will be used.
6. Basic performance of the equipment and the system
Care Optoelectronics
Corporaon │ X-
Ray Products
B.2 Further Information for IEC
2 (EN 60601
-1-2)
_____________________________________________________________
This equipment is not intended for use in residential environments and may not
provide adequate protection
to radio reception in such environments.
_____________________________________________________________
1. Medical electrical equipment needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information
rovided in the accompanying documents.
2. Portable and mobile RF communications equipment can affect medical
3. Information regarding the cable affecting EMC is as follows.
Connection
Maximum
length
General specification
Extend Back
-up
cable to PC or
network switch.
20m
Cat 5e or more
UTP type and straight
cable
From adaptor to AC
Use a hospital grade power cord
4. The use of accessories, transducers and cables other than those specified,
e exception of transducers and cables sold by
InnoCare Optoelectronics
as replacement parts for internal components, may result in increased
emissions or decreased immunity of the Yushan.
d not be used adjacent to or stacked with o
If adjacent or stacked use is necessary, the Yushan FPD should be observed
to verify normal operation in the configuration in which it will be used.
6. Basic performance of the equipment and the system
71
Ray Products
_____________________________________________________________
This equipment is not intended for use in residential environments and may not
to radio reception in such environments.
_____________________________________________________________
1. Medical electrical equipment needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information
2. Portable and mobile RF communications equipment can affect medical
General specification
Cat 5e or more
UTP type and straight
Use a hospital grade power cord
4. The use of accessories, transducers and cables other than those specified,
InnoCare Optoelectronics
as replacement parts for internal components, may result in increased
d not be used adjacent to or stacked with o
ther
If adjacent or stacked use is necessary, the Yushan FPD should be observed
to verify normal operation in the configuration in which it will be used.