User Manual
65
© InnoCare Optoelectronics Corporaon │ X-Ray Products
image quality and electro-optical performances are not guaranteed. It also
means that, except if mentioned explicitly, any perturbation occurring within
those conditions are reversible and not existing anymore when the detector is
turned back within the performance operating conditions.
Item Min Typ. Max
Temperature
5°C 25°C 35°C
Humidity 15% 80%
Pressure 700mBar 1013mBar 1060mBar
Weight (uniform load) 300 kg
Weight (local load on 4 cm
diameter)
120 kg
Vibrations (10 -200 Hz / 3
axis )
2 g
Drop height (the reversibility
to full performance is not
guaranteed for this item)
70 cm
A.1.5.3 Performance Operating
Performance is guaranteed in the operating range below.
Item Min. Typ. Max.
Temperature
15°C 25°C 30°C
Humidity 15% 80%
Pressure 700mBar 1013mBar 1060mBar
Distance to calibration
temperature
-12°C 0°C +12°C
Weight (uniform load) 300 kg
Weight (local load on 4 cm
diameter disc)
120 kg
Vibrations (10 -200 Hz / 3
axis )
2 g
Drop (The Yushan
V14C/V14G/V17C/V17G
remains at full performance
after a drop test of following
height)
70 cm
_____________________________________________________________