Chapter 6 Troubleshooting
- In case of smoke, fire, abnormal high temperature, remove battery and
power supply and turn off detector imediately.
- In case of yellow error LED on, it may trigger the impact. Please read the
manual to release the status and restart the power; if the yellow light
cannot be removed, contact InnoCare service engineer or local
representative
- In case of abnormal image, calibrate detector by standard method.
Contact InnoCare service engineer or local representative if error is not
removed by calibration.
- Other technical issue, contact InnoCare service engineer or local
representative.
- When Yushan cannot be turned on, please check if the battery has power
or if the DC power supply is connected.
- When the software shows that the network can't connect, please confirm
the signal of Yushan first. If it is wireless network, the third light needs to
be solid green to indicate that it has been connected. If the green light
flashes, it means that Yushan is not connected. At this time, please check
the network settings of the environment and the computer, and have
successfully connected.