4.5.4 Standard Configuration
A. Other Essential System Components necessary to perform Digital
Radiography
Yushan (Supporting SDK Interface) can be used in combination with the
image processing unit provided by System Integrator.
The X-ray equipment is composed of X-ray high voltage generator and
X-ray console, etc. The hand switch is connected with the console for the
exposure. At first, when making an exposure, press the Prep switch only
of the hand switch. After confirming that the X-ray equipment is ready
System Integrator of X-rays, press the Exp. switch of the hand switch.
The Access Point is required system, when Yushan is used in wireless
communication mode.
B. A standard configuration of this device is as below.
Yushan (Supporting SDK Interface):
Flat panel sensor : Yushan V14C / V14G / V17C / V17G
The control software is installed and used in a Windows PC (the image
processing unit) prepared by the System Integrator.
This variation provides images, which diagnostic image process is not
implemented, to the image processing unit.
An X-ray automatic detection function is used for Yushan V14C / V14G / V17C
/ V17G. This function could detect a start timing of an X-ray exposure using
readout signals in the detector. After detecting the start timing of the X-ray
exposure, the detector starts collecting of X-ray signals and image readout.