4.3.2 Overview of User Interface
Operation
o Enter Operation: Make FPD enter operation state. User can choose exposure mode
after FPD is in operation state.
o Exit Operation: Make FPD exit operation state.
o Set FPD time: Synchronize the time between PC and RTC of FPD.
Mode
o Start: Make FPD operate in selected exposure mode.
o Stop: Make FPD stop selected exposure mode and return to Idle in operation state.
o Reset: Reset the image module (Clear some alert flag).
o Mode: Choose exposure mode (AED repeat / AED one shot / Soft Sync /
Acquisition / Discharge).
o Auto Download: Download the image to PC after the exposure.
o Don’t Save to eMMC:
▪☑ The image is saved in the non-volatile storage of FPD.
▪☐ The image is only stored in buffer of FPD and will be overwritten by
subsequent exposure image.
o Corrected Image: Download the corrected image of AED mode. (App ignores this
parameter if FPD is not operated in AED mode)
o User Key: A number for user key generation. The number increases by 1 after each
exposure.
Exposure Configuration (Please refer to AED / Acquisition section for more
detail)
o Offset:
▪☑ Taking offset image depends on exposure mode
• Offset After Exposure (OAX): AED
• Offset Before Exposure (OBX): Acquisition
▪Do not take offset image
o Clear1: Discharge before take OBX of Acquisition mode.
o Delay OBX: If this option is checked, there will be an x-ray window delay between
discharge and OBX. Otherwise the OBX is acquired immediately after Clear1.
o Clear2: Discharge before take exposure image.
o X-Ray Window (ms): X-Ray window in milliseconds. [2.5ms~3s]
o Offset Window (ms): Offset window in milliseconds. [2.5ms~3s]