Operators Manual Instruction Manual

800-446-7488 37 www.in-situ.com
Problem Possible Solution
Unable to connect the TROLL
Link Telemetry System to Win-
Situ 5 Plus Software in TCP/IP
mode.
Ensure the correct IP Address is entered. See page
24.
Ensure the correct Port Number is entered. See page
24.
Ensure the correct Device Address is entered. See
page 24.
Ensure the communication mode is set to ASCII. See
page 9.
Ensure the Transmission Delay is set between 15 to
20 seconds. See page 24.
Ensure all communication settings on the instrument
are correct. See page 8.
Ensure the APN information is entered correctly using
the TROLL Link Setup Software. See page 24.
Unable to connect the TROLL
Link Telemetry System to Win-
Situ 5 Plus Software in dial-up
mode.
Ensure the correct Modem number is selected. See
page 23.
Ensure the correct phone number is entered. See
page 23.
Ensure the correct Device Address is entered.
See page 23.
Ensure the communication mode is set to ASCII. See
page 9.
Ensure the Transmission Delay is set between 15 to
20 seconds. See page 23.
Ensure all communication settings on the instrument
are correct. See page 8.
Unable to connect the TROLL
Link Telemetry System to Win-
Situ 5 Plus Software in SMS
mode.
A different mode must be selected when a direct
connection to the system is required.