Operators Manual
39
8. Rinse the sensors with DI water.
Calibrate the Rugged Dissolved Oxygen Sensor
The RDO sensor requires very little maintenance. The zero oxygen calibration is
optional.
1. Tap the Calibration icon .
2. Tap RDO Sensor.
3. Select the method by which you intend to calibrate the sensor. This example
demonstrates a two-point calibration. Tap 100% and 0% Saturation.