Application
TheRDO®Sensorisextremelystable,whicheliminates
theneedforfrequentcalibration.Standardoperating
procedures,environmentalconditions(e.g.,potential
forfouling,abrasiveconditions)maynecessitate
routinecalibration.
AfterinstallinganewRDOSensor,In-Situ®recommends
thatyouperformacalibration.Forbestresults,perform
a2-pointcalibrationat100%and0%dissolvedoxygen
(DO)saturationbeforeusingthesensor.
100% Oxygen Saturation Calibration
1. Installthebatteriesintheportableairpumpas
showninthediagramontheinsidecover.
2. Fillthelargecalibrationcuptowithin4cm(1.5in.)
ofthetopwithcleanwater.
Note: Use tap water, distilled water, or your sample matrix if it
is freshwater, not brackish or saline.
3. Turnonthebubbler,openthepinchvalveonthe
tubing.Runatfullowratefor10minutestofully
saturatethewaterwithoxygen.After10minutes
youcanusethecontrolvalvetomatchthewater
turbulenceconditionsexpectedintheeld.
4. Placetheinstrumentinthewaterensuringthat
thesensorandthetemperaturethermistorare
completelysubmergedandthattheRDOsensing
materialisnotinthedirectpathofthebubbles.
5. Connecttheinstrumenttoacomputerand
establishaconnectiontothecalibrationsoftware.
6. Followthecalibrationwizardprovidedinthesoftware.
Note: Video tutorials and operators manuals are available
on the In-Situ website www.in-situ.com in the Tech
Resources section.
0% Oxygen Saturation Calibration
1. Fillthesmallercalibrationcuptotheupperllline
withsodiumsulte(NA
2
SO
3
)solution.
2. Placetheinstrumentinthesolutionandallowtime
fortemperaturestabilization.
3. Ensuretherearenoairbubblesonthesensing
surfaceoftheRDOSensor.
4. Followthecalibrationwizardprovidedinthesoftware.
Note: Video tutorials and operators manuals are available
on the In-Situ website www.in-situ.com in the Tech
Resources section.
5. Whenyouarenishedrinsethoroughlyincleanwater.
Replacement Parts
Sodium sulte solution, 500 mL 0017670