Manual
1 800 446 7488
(toll-free, US and Canada) or 970 498 1500 www.in-situ.com
Multi-Parameter Water Quality TROLL
®
This information is subject to change without notice. In-SituandtheIn-Situlogo,Win-Situ,Pocket-Situ,TROLL,RuggedCable,RuggedReader,
and RDO are trademarks or registered trademarks of In-Situ Inc. Teflon is a registered trademark of E.I. DuPont de Nemours and Company.
Copyright © 2002 – 2010 by In-Situ Inc. All rights reserved.
0044712 rev. 002 04/10
5. Attach hoses to valves and secure with clamps (may not be
needed with very low flow rates).
6. Toprovidestability,attachspiketobaseofowcellandpress
into the ground. Can be secured at a small angle from vertical
(< 30°) to allow bubbles to self-purge from the cell.
7. Remove the restrictor and insert the calibrated MP TROLL into
the cylinder. Tighten threads to attach (not too tight).
physiCAL DesCription
Dimensions 6.4cm(2.5in)O.D.,21.4cm
(8.4 in) long
Volume 200 mL without sensors
Materials Clear polished polycarbonate
Groundspike 316Lstainlesssteel
NPT attachment
3
/8” NPT
Hose sizes accepted
3
/8”,½”
Typical flow rate 0.1 - 0.5 liters/minute
Recommended Flow rate: 8.0 liters/minute
operational limits Pressure: 30 psi
ACCessories CAtALog no.
Flatbase12"x12"PVC,withhardware .................... 44430
Flow-Sense™ low-flow software .....................................NA
(included in Win-Situ 4 and Pocket-Situ 4, version 4.40 and later)
Hardware fittings kit ................................................... 44340
(contains 4 barbed adapters, 6 hose clamps, pipe thread tape)
guiDeLines AnD preCAutions
• Flowiscontrolledbythepump,notthe3-wayballvalve.The
valve may be used to divert flow for sample collection.
Low -FLow CeLL
Catalog No. 0044710
• Todecreaseairbubblesinthesystem,tapetheadapterttings
before installation.
• Topreventinterferencewithsensors,ensurethatadaptert-
tings do not protrude beyond the inside wall of the flow cell.
Replacement
3
/8” NPT adapter fittings are available from In-Situ
Inc. readily available at hardware stores and plumbing supply
houses.
• In-Situ'sFlow-Senselow-owsoftwarecalculatesvolumeauto-
matically.
• Cleananddecontaminatetheinnerstainlesssteelsleeveafter
use.Re-install,eithersideup.
to pump
Check valve
Ball valve
Optional
turbidity
sleeve
Spike
(or base plate)
←Flow